The possibility to rationalize and simplify bioanalysis, without compromising the analytical quality, by omitting the calibration curves was studied. Using mass spectrometry (MS) and a stable isotope labeled internal standard it was possible to get equally good results by calculating the results directly from the analyte/internal standard area ratio and a predetermined response factor as by the traditional way, using a calibration curve run at the same occasion. To be able to use this simplified quantification method, that we call internal calibration, in its most simple form there are some prerequisites that must be considered: (1) The relative response should not be concentration dependent. (2) The relative response should be constant between batches/days. (3) The level of analyte in the internal standard should not be detectable. (4) There should be no influence from naturally occurring isotopes of the analyte on the internal standard peak area. A bioanalytical LC-MS/MS method for a research compound was validated both with and without calibration curves and no significant differences were found regarding precision and accuracy. It was shown that all four prerequisites above were fulfilled. Validation data were very good for the whole concentration range, 0.010-30 micromol/L. Long-term data for QC samples showed excellent precision and accuracy.
机译:研究了通过省略校准曲线来合理化和简化生物分析而不损害分析质量的可能性。使用质谱(MS)和稳定的同位素标记的内标,可以通过使用传统方法通过使用校准曲线运行直接从分析物/内标面积比和预定响应因子计算结果,获得同样好的结果在同一场合。为了能够使用这种简化的量化方法(我们称为内部校准),必须以其最简单的形式来考虑一些先决条件:(1)相对响应不应与浓度有关。 (2)批次/天之间的相对响应应恒定。 (3)内标中的分析物含量应不可检测。 (4)分析物的天然同位素对内标峰面积没有影响。使用和不使用校准曲线均验证了用于研究化合物的生物分析LC-MS / MS方法,并且在精密度和准确度上均未发现明显差异。结果表明,以上所有四个先决条件都已满足。整个浓度范围为0.010-30 micromol / L的验证数据都非常好。质控样品的长期数据显示出极好的精度和准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号