首页> 外文期刊>Rapid Communications in Mass Spectrometry: RCM >Phospholipids in liquid chromatography/mass spectrometry bioanalysis: comparison of three tandem mass spectrometric techniques for monitoring plasma phospholipids, the effect of mobile phase composition on phospholipids elution and the association of phospholipids with matrix effects
【24h】

Phospholipids in liquid chromatography/mass spectrometry bioanalysis: comparison of three tandem mass spectrometric techniques for monitoring plasma phospholipids, the effect of mobile phase composition on phospholipids elution and the association of phospholipids with matrix effects

机译:液相色谱/质谱生物分析中的磷脂:监测血浆磷脂的三种串联质谱技术,流动相成分对磷脂洗脱的影响以及磷脂与基质效应的关系

获取原文
获取原文并翻译 | 示例
           

摘要

Because plasma phospholipids may cause matrix effects in bioanalytical liquid chromatography/tandem mass spectrometry (LC/MS/MS) methods, it is important to establish optimal mass spectrometric techniques to monitor the fate of phospholipids during method development and application. We evaluated three MS/MS techniques to monitor phospholipids using positive and negative electrospray ionization (ESI). The first technique is based on using positive precursor ion scan of m/z 184, positive neutral loss scan of 141 Da and negative precursor ion scan of m/z 153. The second technique is based on using class-specific positive and negative selected reaction monitoring (SRM) transitions to monitor class-representative phospholipids. The third technique, previously reported, utilizes in-source collision-induced dissociation (CID)-based positive SRM of m/z 184 -> 184. We recommend the all-inclusive technique 1 for use in qualitative assessment of all classes of phospholipids and technique 2 for use in quantitative assessment of class-representative phospholipids. Secondly, we evaluated the elution behaviors of the plasma phospholipids under different reversed-phase mobile phase conditions. The phospholipid-eluting strength of a mobile phase was mainly dependent on the type and amount M of the organic eluent and the strength increased in the order of methanol, acetonitrile and isopropyl alcohol. Under the commonly used gradient and isocratic elution schemes in LC/MS/MS bioanalysis, not all the phospholipids are eluted off the column. Thirdly, we investigated the association between phospholipids and matrix effects in positive and negative EST using basic, acidic and neutral analytes. While the phospholipids caused matrix effects in both positive and negative EST, the extent of ionization suppression was analyte-dependent and was inversely related to the retention factor and broadness of the phospholipids peaks. The lysophospholipids which normally elute earlier in reversed-phase chromatography are more likely to cause matrix effects compared to the later-eluting phospholipids in spite of the larger concentrations of the latter in plasma.
机译:由于血浆磷脂可能会在生物分析液相色谱/串联质谱(LC / MS / MS)方法中引起基质效应,因此建立最佳质谱技术以监测方法开发和应用过程中磷脂的命运非常重要。我们评估了三种使用正电和负电喷雾电离(ESI)监测磷脂的MS / MS技术。第一种技术基于使用m / z 184的正前体离子扫描,141 Da的正中性损失扫描和m / z 153的负前体离子扫描。第二种技术基于使用特定于类别的正负选择反应监测(SRM)转换以监测类代表的磷脂。先前报道的第三种技术利用了基于源碰撞诱导解离(CID)的m / z 184-> 184的正SRM。我们建议采用全包技术1对所有类别的磷脂和脂质进行定性评估。技术2用于定量评估类代表性磷脂。其次,我们评估了在不同的反相流动相条件下血浆磷脂的洗脱行为。流动相的磷脂洗脱强度主要取决于有机洗脱液的类型和量M,强度按甲醇,乙腈和异丙醇的顺序增加。在LC / MS / MS生物分析中常用的梯度和等度洗脱方案下,并非所有的磷脂都从色谱柱上洗脱下来。第三,我们使用碱性,酸性和中性分析物研究了正,负EST中磷脂与基质效应之间的关系。尽管磷脂在正和负EST中均引起基质效应,但电离抑制的程度却取决于分析物,并且与保留因子和磷脂峰的宽度成反比。尽管后来血浆中的磷脂浓度较高,但通常在反相色谱中较早洗脱的溶血磷脂与后来洗脱的磷脂相比更可能引起基质效应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号