A high dispersion FIA system with a fluoride selective and a double junction reference electrode in cascade arrangement was developed for the determination of the fluoride content in chromium electroplating baths. The analyses of seven synthetic baths prepared in the laboratory and two samples provided by the industry gave results with a mean recovery value of 100.5%, and a mean variation coefficient of 1.6%. These data were better than those obtained by the EDTA complexometric titration, the reference technique, with a mean recovery of 94.0% and a mean variation coefficient of 7.5%. The sampling rate for the FIA manifold was 20 samples per hour. [References: 17]
机译:开发了一种具有氟选择性电极和双级参比电极级联排列的高分散FIA系统,用于测定铬电镀液中的氟含量。对实验室中制备的7种合成浴液和工业界提供的2种样品进行分析后得出的结果是,平均回收率为100.5%,平均变异系数为1.6%。这些数据优于通过EDTA络合滴定法(参考技术)获得的数据,平均回收率为94.0%,平均变异系数为7.5%。 FIA歧管的采样率为每小时20个样本。 [参考:17]