首页> 外文期刊>Analytical Biochemistry: An International Journal of Analytical and Preparative Methods >Comparison of ultrafiltration units for proteomic and N-glycoproteomic analysis by the filter-aided sample preparation method
【24h】

Comparison of ultrafiltration units for proteomic and N-glycoproteomic analysis by the filter-aided sample preparation method

机译:过滤辅助样品制备方法用于蛋白质组学和N-糖蛋白组学分析的超滤单元的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The filter-aided sample preparation (FASP) method allows gel-free processing of biological samples solubilized with detergents for proteomic analysis by mass spectrometry. In FASP detergents are removed by ultrafiltration, and after protein digestion peptides are separated from undigested material. Here we compare the effectiveness of different filtration devices for analysis of proteomes and glycoproteomes. We show that Microcon and Vivacon filtration units with nominal molecular weight cutoffs of 30,000 and 50,000 (30 and 50 k, respectively) are equally suitable for FASP, whereas Microcon 30 k units are most appropriate for mapping of N-glycosylation sites. The use of filters with these relatively large cutoffs facilitates depletion of detergents.
机译:借助过滤器辅助的样品制备(FASP)方法,可以无胶处理溶解有去污剂的生物样品,从而通过质谱进行蛋白质组学分析。在FASP中,通过超滤除去去污剂,然后在蛋白质消化后将肽与未消化的物质分离。在这里,我们比较了用于蛋白质组学和糖蛋白组学分析的不同过滤装置的有效性。我们显示标称截留分子量为30,000和50,000(分别为30和50 k)的Microcon和Vivacon过滤装置同样适用于FASP,而Microcon 30 k单位最适合于N-糖基化位点的定位。具有这些较大截止值的过滤器的使用促进了洗涤剂的消耗。