首页> 外文期刊>Journal of Radioanalytical and Nuclear Chemistry: An International Journal Dealing with All Aspects and Applications of Nuclear Chemistry >Alpha spectroscopy backgrounds at Lawrence Livermore National Laboratory: technical equivalency for backgrounds with blank plates versus backgrounds with no plates
【24h】

Alpha spectroscopy backgrounds at Lawrence Livermore National Laboratory: technical equivalency for backgrounds with blank plates versus backgrounds with no plates

机译:alpha光谱背景在Lawrence Livermore国家实验室:背景的技术等效与空白板的背景与没有盘子的背景

获取原文
获取原文并翻译 | 示例
           

摘要

Alpha spectroscopy chamber backgrounds are routinely measured using blank plates. Operator time spent moving blank plates in and out of chambers is significant. If chamber backgrounds are measured in empty chambers, time spent moving blank plates can be reclaimed for other uses. This study compares results of chamber backgrounds measured with and without plates. Five hundred fourteen chamber backgrounds measured with plates are paired with 514 chamber backgrounds measured without plates. The total counts within 13 spectral regions of interest are compared statistically. Chamber backgrounds measured without blank plates are found statistically equivalent to chamber backgrounds measured with blank plates.
机译:alpha光谱学室背景使用空白板常规测量。 操作员时间花空白板进出腔室是显着的。 如果在空腔室中测量腔室背景,则可以为其他用途回收时间支出移动空白板。 该研究比较了用腔室测量的腔室背景的结果。 用板材测量的五十四个室背景与无板材测量的514室背景配对。 在统计上比较13个光谱区域内的总计数。 没有空白板测量的室背景统计上相当于用空白板测量的腔室背景。