Comparing optic nerve head analysis between confocal scanning laser ophthalmoscopy and spectral domain optical coherence tomography

机译:共焦扫描激光检眼镜和光谱域光学相干断层扫描技术之间的视神经头部分析比较

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose/aim of the study: Confocal scanning laser ophthalmoscopy, HRT3, and spectral domain optical coherence tomography (OCT), RTVue-100, are able to give 3-dimensional (3D) topography images of optic nerve head (ONH) and to derive stereometric parameters and sectorial analysis. The purpose of the study is to evaluate the agreement of these two devices and their diagnostic accuracy to discriminate eyes with glaucoma from those without.Materials and methods: Glaucoma patients and healthy control subjects were included. All of them underwent a complete ophthalmological examination, including slit lamp evaluation and visual field (VF) test. After pupil dilatation, HRT3 and RTVue-100 were performed. The following stereometric parameters were recorded: disc area, rim area, rim volume, cup volume, cup area, cup/disk ratio, and the following sectors, superotemporal, superonasal, inferotemporal, inferonasal.Results: Forty-six eyes of 46 glaucoma patients and 58 eyes of 58 healthy subjects were included in the study. In both groups, HRT3 rim area and rim volume were statistically higher than RTVue-100 (glaucomas: 0.95 ± 0.38 versus 0.44 ± 0.33 and 0.19 ± 0.13 versus 0.02 ± 0.03, p < 0.01. controls: 1.41 ± 0.30 versus 1.08 ± 0.37 and 0.37 ± 0.13 versus 0.14 ± 0.11, p < 0.01), while cup area was statistically higher by RTVue-100 (glaucomas: 1.42 ± 0.57 versus 1.14 ± 0.58, p < 0.01. controls: 1.05 ± 1.35 versus 0.65 ± 0.48). Bland and Altman plots confirmed the presence of a fixed bias. The parameters with largest AUROC were rim volume, rim area and cup/disk ratio for both instruments. HRT3 inferotemporal sector had the highest sensitivity (80.43%, at 75.9% specificity), while for RTVue-100, the superotemporal sector had the highest sensitivity (76.1%, at 81% specificity). The agreement was moderate for inferotemporal sector and fair for the others.Conclusions: HRT3 and RTVue-100 are not interchangeable for ONH analysis. They both have good diagnostic accuracy, but RTVue-100 shows slightly better performance, at least with regard to rim area volume.
机译:研究目的/目的:共聚焦扫描激光检眼镜HRT3和光谱域光学相干断层扫描(OCT)RTVue-100,能够给出视神经头(ONH)的3维(3D)地形图并得出立体参数和部门分析。该研究的目的是评估这两种设备的一致性及其诊断准确性,以区分有无青光眼和无青光眼。材料与方法:包括青光眼患者和健康对照组。他们都接受了完整的眼科检查,包括裂隙灯评估和视野(VF)测试。瞳孔扩张后,进行HRT3和RTVue-100。记录以下立体参数:椎间盘面积,轮缘面积,轮缘体积,杯体积,杯面积,杯/盘比率以及以下扇形,颞上,鼻上,颞下,鼻下。结果:46名青光眼患者的46眼研究包括58位健康受试者的58只眼睛。两组的HRT3轮辋面积和轮辋体积在统计学上均高于RTVue-100(青光眼:0.95±0.38对0.44±0.33和0.19±0.13对0.02±0.03,p <0.01。对照组:1.41±0.30对1.08±0.37和0.37±0.13对0.14±0.11,p <0.01),而RTVue-100的杯面积在统计学上更高(青光眼:1.42±0.57对1.14±0.58,p <0.01。对照:1.05±1.35对0.65±0.48)。布兰德和奥特曼的图证实存在固定偏差。 AUROC最大的参数是两种仪器的轮辋体积,轮辋面积和杯盘比。 HRT3颞下部门具有最高的敏感性(80.43%,特异性为75.9%),而RTVue-100的颞下部门具有最高的敏感性(76.1%,特异性为81%)。该协议对下颞部领域是中等的,对其他领域是公平的。结论:HRT3和RTVue-100不可互换用于ONH分析。它们都具有良好的诊断准确性,但是RTVue-100的性能至少在边缘区域体积方面显示出更好的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号