首页> 外文期刊>Medical Physics >SU‐F‐T‐76: Total Skin Electron Therapy: An‐End‐To‐End Examination of the Absolute Dosimetry with a Rando Phantom
【24h】

SU‐F‐T‐76: Total Skin Electron Therapy: An‐End‐To‐End Examination of the Absolute Dosimetry with a Rando Phantom

机译:SU-F-T-76:总皮肤疗法:与Rando Phantom的绝对剂量测定的端到端检查

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To examine and validate the absolute dose for total skin electron therapy (TSET) through an end‐to‐end test with a Rando phantom using optically stimulated luminescent dosimeters (OSLDs) and EBT3 radiochromic films. Methods: A Varian Trilogy linear accelerator equipped with the special procedure 6 MeV HDTSe‐ was used to perform TSET irradiations using a modified Stanford 6‐dual‐field technique. The absolute dose was calibrated using a Markus ion chamber at a reference depth of 1.3cm at 100 cm SSD with a field size of 36 × 36 cm at the isocenter in solid water slabs. The absolute dose was cross validated by a farmer ion chamber. Then the dose rate in the unit of cGy/Mu was calibrated using the Markus chamber at the treatment position. OSLDs were used to independently verify the dose using the calibrated dose rate. Finally, a patient treatment plan (200 cGy/cycle) was delivered in the QA mode to a Rando phantom, which had 16 pairs of OSLDs and EBT3 films taped onto its surface at different anatomical positions. The doses recorded were read out to validate the absolute dosimetry for TSET. Results: The OSLD measurements were within 7% agreement with the planned dose except the shoulder areas, where the doses recorded were 23% lower on average than those of the planned. The EBT3 film measurements were within 10% agreement with the planned dose except the shoulder and the scalp vertex areas, where the respective doses recorded were 18% and 14% lower on average than those of the planned. The OSLDs gave more consistent dose measurements than those of the EBT3 films. Conclusion: The absolute dosimetry for TSET was validated by an end‐to‐end test with a Rando phantom using the OSLDs and EBT3 films. The beam calibration and monitor unit calculations were confirmed.
机译:目的:使用光学刺激的发光剂量计(OSLD)和EBT3放射晶片,通过与RANDO Phantom的端到端测试检查和验证总皮肤疗法(TSet)的绝对剂量。方法:使用改进的斯坦福6-Dual-Field技术,使用配备有特殊程序的Varian Trilogy线性加速器6 Mev HDTSE-用于执行TSet照射。使用Markus离子室在1.3cm的参考深度下在100cm SSD的参考深度处使用Markus离子室校准绝对剂量,其在固体水平板中的Isocenter处的场尺寸为36×36cm。通过农民离子室验证的绝对剂量。然后使用Markus腔室在治疗位置校准CGY / MU单元中的剂量率。 OSLD用于独立地使用校准剂量率验证剂量。最后,将患者治疗计划(200CGY /循环)以QA模式输送到RANDO Phantom,其在不同的解剖位置以16对OSLD和EBT3薄膜贴在其表面上。记录的剂量被读出以验证TSet的绝对剂量剂。结果:OSLD测量与除肩部区域以外的计划剂量达到7%,平均记录的剂量降低23%。 EBT3薄膜测量在与肩部和头皮顶点区域以外的计划剂量达成10%的协议,而记录的各个剂量的平均值比计划的相应剂量为18%和14%。 OSLDS比EBT3薄膜更一致的剂量测量。结论:使用OSLD和EOBT3薄膜的rando Phantom,通过端到端测试验证了Tset的绝对剂量测定。梁校准和监测单元计算得到确认。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号