首页> 外文期刊>Journal of magnetic resonance >4 T split TEM volume head and knee coils for improved sensitivity and patient accessibility
【24h】

4 T split TEM volume head and knee coils for improved sensitivity and patient accessibility

机译:4 T分体式TEM头和膝线圈,提高了灵敏度和患者可及性

获取原文
获取原文并翻译 | 示例
           

摘要

Split RF coils offer improved patient access by eliminating the need for the coil to be slid over the region of interest. For unshielded birdcage coils, the presence of end ring currents necessitates a direct electrical connection between two halves of the coil. For high-field (> 3 T) shielded birdcage coils, both the shield and the coil must be split and reliably connected electrically. This problem can be circumvented by the use of split TEM volume coils. Since the elements of a TEM coil are coupled inductively, no direct electrical connection between the halves is necessary. In this work we demonstrate that the effects of splitting the shield for head and knee TEMs can be compensated for, and performance retained. For the knee, the improved access allowed the coil diameter to be reduced, enhancing the sensitivity by 15-20%. (c) 2007 Elsevier Inc. All rights reserved.
机译:分离式RF线圈消除了在感兴趣区域上滑动线圈的需要,从而改善了患者的出入通道。对于非屏蔽鸟笼线圈,端环电流的存在需要在线圈的两半之间直接进行电连接。对于高磁场(> 3 T)的屏蔽鸟笼线圈,屏蔽层和线圈都必须分开并可靠地电气连接。可以通过使用分体式TEM体积线圈来解决此问题。由于TEM线圈的元件是电感耦合的,因此两半之间无需直接电连接。在这项工作中,我们证明了头部和膝盖TEM屏蔽层的分割效果可以得到补偿,并且性能得以保持。对于膝盖而言,改善的通道允许减小线圈直径,从而将灵敏度提高15-20%。 (c)2007 Elsevier Inc.保留所有权利。