Free forms of light chains (FLCs) in serum are clinically useful markers, but their practical use in laboratories has been limited due to the cross-reaction with light chains associated with heavy chains (intact immunoglobulins) on FLC assays. In this study, we attempted to use anti-FLC antiserum and anti-FLC monoclonal antibody (mAb) in a sandwich enzyme-linked immunosorbent assay (ELISA) for serum FLCs. This combination resulted in a synergistic effect, showing 10(5)-fold specificity for FLC kappa and 10(4)-fold for FLC lambda compared to IgG. The specific ELISAs had good sensitivity, detecting 0.78 mug/l of FLC kappa and 31.3 mug/l of FLC lambda. We then measured serum samples before and after the absorption of IgG, IgA, and IgM and found that there was no appreciable difference in the FLC values between them. These results indicate that the ELISAs quantify serum FLCs more precisely even in the presence of about a thousand-fold amount of intact immunoglobulin. Meanwhile, the use of two anti-FLC mAbsin an ELISA showed no response for FLCs probably because of the conflict of the two mAbs at the specific epitope in FLC. Thus, ELISAs with anti-FLC antiserum and anti-FLC mAb can provide a reliable method for the clinical test of serum FLCs.
机译:血清中游离形式的轻链(FLC)是临床有用的标记物,但是由于FLC分析中与重链相关的轻链(完整免疫球蛋白)的交叉反应,它们在实验室中的实际应用受到了限制。在这项研究中,我们尝试在抗夹心酶联免疫吸附法(ELISA)中对血清FLC使用抗FLC抗血清和抗FLC单克隆抗体(mAb)。这种组合产生了协同效应,与IgG相比,对FLCκ的特异性为10(5)倍,对FLCλ的特异性为10(4)倍。特异性ELISA具有良好的灵敏度,可检测到0.78杯/升的FLCκ和31.3杯/升的FLCλ。然后,我们在吸收IgG,IgA和IgM之前和之后测量了血清样品,发现它们之间的FLC值之间没有明显的差异。这些结果表明,即使存在约一千倍的完整免疫球蛋白,ELISA也可以更精确地定量血清FLC。同时,在ELISA中使用两种抗FLC单克隆抗体对FLC没有反应,这可能是由于两种单克隆抗体在FLC中特定表位的冲突所致。因此,具有抗FLC抗血清和抗FLC mAb的ELISA可以为血清FLC的临床测试提供可靠的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号