首页> 外文期刊>Journal of Electron Microscopy >Energy-drift correction of electron energy-loss spectra from prolonged data accumulation of low SNR signals.
【24h】

Energy-drift correction of electron energy-loss spectra from prolonged data accumulation of low SNR signals.

机译:通过低SNR信号的长时间数据累积,对电子能量损失谱进行能量漂移校正。

获取原文
获取原文并翻译 | 示例
           

摘要

The demand for analysis of trace elements in atomically localized areas by electron energy-loss spectroscopy (EELS) in transmission electron microscopy is increasing. Unfortunately, the prolonged data acquisitions required to achieve an acceptable signal-to-noise ratio (SNR) tend to deteriorate the energy resolution because of spectral drifts due to instrumental instability. We developed a macroscript for a Gatan Digital Micrograph(TM) to control an ENFINA 1000 EEL spectrometer that corrects for energy drifts during data accumulation. The script successfully achieved a core-loss spectrum for a sample having approximately 1 at% elemental concentration, and provided a sufficient SNR for chemical state analysis.
机译:透射电子显微镜中通过电子能量损失谱(EELS)分析原子局部区域中的痕量元素的需求正在增加。不幸的是,由于仪器的不稳定性,频谱漂移会导致获得可接受的信噪比(SNR)所需的长时间数据采集趋于降低能量分辨率。我们为Gatan Digital Micrograph(TM)开发了一个宏脚本,以控制ENFINA 1000 EEL光谱仪来校正数据累积过程中的能量漂移。该脚本成功实现了元素浓度​​约为1 at%的样品的核心损耗光谱,并为化学状态分析提供了足够的SNR。