首页> 外文期刊>Journal of endovascular therapy: an official journal of the International Society of Endovascular Specialists >Staged proximal deployment of the Zenith TX2 thoracic stent-graft: a novel technique to improve conformance to the aortic arch.
【24h】

Staged proximal deployment of the Zenith TX2 thoracic stent-graft: a novel technique to improve conformance to the aortic arch.

机译:Zenith TX2胸腔支架移植物的分阶段近端部署:改善对主动脉弓顺应性的新技术。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To present a modification of the Zenith TX2 thoracic stent-graft that allows staged proximal deployment to improve apposition to the aortic wall. TECHNIQUE: Three standard Zenith TX2 thoracic stent-grafts and 3 modified versions were deployed in a glass model of the aortic arch. Deployment sequences were analyzed. In a patient with a 6-cm thoracic aortic aneurysm after a type B dissection, the modified Zenith TX2 thoracic stent-graft was deployed successfully and without complications; the proximal part of the stent-graft protruded less into the arch, significantly improving wall apposition. CONCLUSION: Staged proximal deployment with a modified Zenith TX2 thoracic stent-graft can improve orientation and wall apposition of the first Z stent. A better proximal apposition may prevent early and late stent-graft complications.
机译:目的:介绍Zenith TX2胸廓支架移植物的一种改良形式,允许分阶段的近端部署以改善与主动脉壁的并置。技术:在主动脉弓的玻璃模型中部署了三个标准的Zenith TX2胸腔支架植入物和3个修改版本。分析了部署顺序。在B型剥离后有6厘米胸主动脉瘤的患者中,改良的Zenith TX2胸支架植入物成功部署且没有并发症。支架移植物的近端较少伸入足弓,显着改善了壁并置。结论:采用改良的Zenith TX2胸腔支架移植物进行近端分阶段部署可以改善第一个Z支架的方向和壁并置。更好的近端并置可以预防早期和晚期支架植入物并发症。