Effect of signal strength on reproducibility of circumpapillary retinal nerve fiber layer thickness measurement and its classification by spectral-domain optical coherence tomography.

机译:信号强度对环乳头视网膜神经纤维层厚度测量可重复性的影响及其光谱域光学相干断层扫描技术的分类。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To assess the effect of signal strength (SS) on the reproducibility of circumpapillary retinal nerve fiber layer (cpRNFL) thickness measurement (measurement agreement) and its color-coded classification (classification agreement) by Cirrus spectral-domain optical coherence tomography (OCT). METHODS: Seven hundred and eighty-nine participants, consecutively enrolled from December 2009 to August 2010, underwent two repeated Cirrus OCT scans to measure cpRNFL thickness (optic disc cube 200 x 200). Intraclass correlation (ICC) and linear-weighted kappa coefficient (kappa) were calculated as indicators for cpRNFL measurement and classification agreement. The difference in cpRNFL thickness measurements between repeated OCT scans (inter-scan measurement difference) was correlated with mean SS, SS change or difference between repeated OCT scans (inter-scan SS difference), and average cpRNFL thickness. RESULTS: Mean SS was negatively correlated with inter-scan SS difference and was positively correlated with the inter-scan measurement difference. Repeated scans with SS change (inter-scan SS difference = 1 or 2) showed larger variability than scans without SS change (inter-scan SS difference = 0) in total average cpRNFL and all quadrant maps except for the nasal quadrant. Multivariate analysis revealed that inter-scan measurement differences increased in subjects with thinner cpRNFL thicknesses and in scans with lower mean SS and higher inter-scan SS differences. Measurement agreement (ICC) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and inferior quadrant maps. Classification agreement (kappa) in groups with larger inter-scan SS differences (=2) was lower than those with smaller inter-scan SS differences (=0 or 1) in their average cpRNFL and quadrant maps (superior and inferior quadrant) although statistically not significant. CONCLUSIONS: Special attention should be paid when comparing the cpRNFL thickness measurement and color-coded classification of 2 or more Cirrus OCT scans as the agreements may be susceptible to SS differences.
机译:目的:通过Cirrus光谱域光学相干层析成像技术(OCT)评估信号强度(SS)对环乳头视网膜神经纤维层(cpRNFL)厚度测量(测量协议)及其色码分类(分类协议)的可重复性的影响)。方法:2009年12月至2010年8月连续招募的789名参与者,进行了两次重复的Cirrus OCT扫描以测量cpRNFL厚度(视盘立方体200 x 200)。计算类内相关性(ICC)和线性加权κ系数(kappa)作为cpRNFL测量和分类一致性的指标。重复OCT扫描之间的cpRNFL厚度测量值的差异(扫描间测量差异)与平均SS,SS变化或重复OCT扫描之间的差异(扫描间SS差异)与平均cpRNFL厚度相关。结果:平均SS与扫描间SS差异呈负相关,与扫描间测量差异呈正相关。重复SS变化(扫描间SS差= 1或2)的扫描显示的总平均cpRNFL和除鼻象限以外的所有象限图的变异性均大于不进行SS变化(扫描间SS差= 0)的扫描。多变量分析显示,在cpRNFL厚度较薄的受试者以及平均SS较低且扫描间SS差异较高的扫描中,扫描间测量差异会增加。在平均cpRNFL和下象限图中,扫描间SS差异较大(= 2)的组的测量一致性(ICC)低于扫描间SS差异较小(= 0或1)的组。尽管在统计学上有统计学差异,但扫描间SS差异较大(= 2)的组的分类一致性(kappa)低于其平均cpRNFL和象限图(上,下象限)中具有较小扫描间SS差异(= 0或1)的组。不重要。结论:比较cpRNFL厚度测量值和2个或更多Cirrus OCT扫描的颜色编码分类时,应格外注意,因为协议可能容易出现SS差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号