BACKGROUND: Left ventricular stimulation for cardiac resynchronization therapy is largely limited by access and anatomy of coronary veins. OBJECTIVE: This study sought to apply ultrasound-mediated leadless pacing technology in heart failure patients and to evaluate the effects of respiration and body posture on the acoustic window. METHODS: Patients with advanced heart failure and ejection fraction
机译:背景:用于心脏再同步治疗的左心室刺激在很大程度上受到冠状静脉的进入和解剖的限制。目的:本研究试图在心力衰竭患者中应用超声介导的无铅起搏技术,并评估呼吸和体位对声学窗的影响。方法:研究了晚期心力衰竭和射血分数<或= 35%的患者。包含接收器电极以进行超声介导的起搏的电生理导管位于左心室。确定了超声介导的起搏阈值。确定患者仰卧时在胸壁上的隔音窗,标题为向左30度,向右30度和向上30度。声学窗口也通过计算机断层扫描和经胸超声心动图确定。用计算机断层摄影术评估了模拟的接收器运动以及呼吸运动和身体位置。结果:对10例患者进行了研究。超声介导的起搏在所有患者中均成功。声学窗口测得为39.6 +/- 18.2 cm(2)。窗口大小随着向右倾斜而减小,而随着向左和向上倾斜而增大。它们与经胸超声心动图和计算机断层扫描的测量结果相关。通过计算机断层扫描估计目标接收器的水平移动水平方向为1.2 +/- 1.4厘米,垂直接收方向为1.3 +/- 0.8厘米。结论:急性心力衰竭患者急性无铅左室刺激的可行性已被证实。经胸部断层超声心动图可预测通过计算机断层扫描验证的声窗。评估呼吸和身体姿势的影响,以开发未来的可植入设备。