To evaluate the effectiveness of the Bipolar Sealer 6.0-VT (BPS 6.0-VT) (Tissue Link Medical, Inc, Dover, NH) in reducing blood loss compared with a conventional electrocautery (Erbe ICC 350) (Erbe ICC 350 (ERBE Elektromedizin GmbH, Tubingen, Germany). A prospective randomized trial of 105 patients with primary total hip arthroplasty (Erbe ICC 350: 50 patients; BPS 6.0-VT: 55 patients). On the basis of the calculation of the preoperative blood volume the total and postoperative blood loss were calculated. No statistically significant difference could be shown for total intraoperative and postoperative blood loss (group A 1846 mL; group B 1740 mL) (t test). Due to the lack of reduction of blood loss and the high costs of the BPS 6.0-VT, its use is not recommended in primary total hip arthroplasty.
机译:与传统的电灼(Erbe ICC 350)(Erbe ICC 350(ERBE Elektromedizin GmbH) ,德国图宾根):一项前瞻性随机试验,共105例原发性全髋关节置换术(Erbe ICC 350:50例; BPS 6.0-VT:55例),根据术前血容量的总和计算失血量,术中和术后总失血量(A组1846 mL; B组1740 mL)(t检验)无统计学意义差异(t检验)。 BPS 6.0-VT,不建议在初次全髋关节置换术中使用。