首页> 美国卫生研究院文献>AJNR: American Journal of Neuroradiology >A Rare Complication of Carotid Artery Stenting: Displacement of Marker Ring Causing Locking of Stent and Incomplete Stent Expansion
【2h】

A Rare Complication of Carotid Artery Stenting: Displacement of Marker Ring Causing Locking of Stent and Incomplete Stent Expansion

机译:颈动脉支架的罕见并发症:标记环的位移导致支架锁定和不完全支架扩张

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

SUMMARY: We present a rare case of a complication of placement of a carotid artery stent represented by partial opening of a carotid Wallstent caused by displacement of its metal ring marker, which thus hindered complete expansion of the stent. An intraluminally locked carotid stent necessitated referral of the patient for urgent carotid endarterectomy. A possible reason of this unusual complication can be a manufacturing defect, which, to our knowledge, was not documented previously in open public data bases or on the Internet.
机译:发明内容:通过其金属环标记物的位移引起的颈动脉壁部分的部分开口呈现颈动脉支架的罕见情况,这是一种罕见的颈动脉支架,从而阻碍了支架的完全膨胀。一个腔内锁定的颈动脉支架需要对患者进行紧急颈动脉胚胎切除术的转介。这种异常并发症的可能原因可以是一种制造缺陷,以我们的知识没有以前在开放的公共数据库或互联网上记录。