Phase-modulation fluorometry using a frequency-doubled pulsed laser diode light source

机译:使用频率加倍脉冲激光二极管光源的相位调制荧光测定法

获取原文

摘要

We used the Hamamatsu model PLP-01 picosecond light pulser as a 413-nm excitation light source for frequency-domain fluorescence measurements. In comparison with sync-pumped/cavity dumped/frequency-doubled dye lasers, the 413-nm PLP-01 shows a longer FWHM (40 ps), a similar pulse repetition rate (up to 10 MHz), much less output power at a fixed wavelength (0.44 mW peak, 220 nW maximum average power), but is less expensive, small-sized, and easy to handle. Using the PLP-01, we have been able to perform fluorescence measurements up to an upper modulation frequency of about 2000 MHz, and to resolve mixtures of fluorophores exhibiting different lifetimes. During our tests, we observed remarkable and lasting (2 h) time drifts between the optical output and the electrical trigger input or output. At present, work is in progress at Hamamatsu to eliminate these drifts.
机译:我们使用Hamamatsu Model PLP-01 PICOSECOND光脉冲器作为413-NM激发光源,用于频域荧光测量。与同步/腔倾倒/倍频染料激光器相比,413-NM PLP-01显示了更长的FWHM(40 ps),类似的脉冲重复率(高达10 MHz),在a下输出功率远低得多固定波长(0.44兆瓦峰值,220兆瓦的最大电源),但不得昂贵,小尺寸,易于处理。使用PLP-01,我们已经能够对大约2000 MHz的上调制频率进行荧光测量,并解析表现出不同寿命的荧光团的混合物。在我们的测试期间,我们观察到光学输出和电触发输入或输出之间的显着和持久(2小时)时间漂移。目前,在滨松的工作正在进行中,以消除这些漂移。