首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Troubleshooting Cross-well Contamination in a 96-Well Plate of Volatile Pharmaceutical Compounds during Evaporation Step using LC/MS/MS
【24h】

Troubleshooting Cross-well Contamination in a 96-Well Plate of Volatile Pharmaceutical Compounds during Evaporation Step using LC/MS/MS

机译:使用LC / MS / MS在蒸发步骤期间挥发性药物化合物的96孔板中的交叉阱污染故障

获取原文

摘要

Changing the analyte ionization state by adjusting the pH of the extracts prior to evaporation, modify the extraction mode or the evaporator model/brand can be very useful to reduce contamination. Lower temperature and minimum N2 gas flow rate should be applied when performing evaporation of small molecules. A new automated liquid handling cross-well contamination test was added in all validation procedures using automated platforms to control such issues. This test is to perform a checkerboard of blanks and high concentration and to alternate for a minimum of 3 consecutive rows on a 96-well plate.
机译:通过在蒸发之前调节提取物的pH来改变分析物电离状态,改变提取模式或蒸发器模型/品牌可以非常有用以减少污染。在进行蒸发小分子时,应施加较低的温度和最小N 2气体流速。使用自动化平台的所有验证程序中添加了一种新的自动化液体处理交叉阱污染测试,以控制这些问题。该测试是执行坯料和高浓度的棋盘,并在96孔板上替换至少3行。