首页> 外文会议>IMEKO TC3 International Conference on Measurement of Force, Mass and Torque >Implementation of the reference force standard in the Croatian metrology and Interlaboratory comparison
【24h】

Implementation of the reference force standard in the Croatian metrology and Interlaboratory comparison

机译:克罗地亚计量与互借比较的参考力标准的实施

获取原文

摘要

Laboratory for Testing Mechanical properties (LIMS) at the Faculty of Mechanical Engineering and Naval Architecture in Zagreb is accredited laboratory for force and a holder of National standard for force in the Republic of Croatia. Reference force standard with measuring range 10 - 500 kN was installed in LIMS at the end of 2005 with the intent of continuous improvement of the metrology infrastructure of Croatia. The load is obtained hydraulically and the measuring system consists of double placed reference transducers with the nominal force 100 kN and 500 kN. Preliminary investigations have shown that the relative measurement uncertainty will be less than 5E-04. Further investigations are current. LIMS has planned to participate in the interlaboratory comparison between two more laboratories. Intercomparison will be done in measuring range 10 - 500 kN using two commercial force transducers of 100 kN and 500 kN nominal values. The best measurements capability (CMC) of the participating laboratories varies from 0,002percent to 0,05percent. The calibration of the force transducers will be carried out using the laboratories own calibration procedure, the relevant classification data, i.e. rotation effect, repeatability, hysteresis as described in DIN EN ISO 376. Intercomparison will increase the level of Croatian metrology thus providing further cooperative investigations in association with European and Global National laboratories for force.
机译:实验室在机械工程和造船的萨格勒布系测试机械性能(LIMS)是认可实验室力和克罗地亚共和国的国家标准为力架。参考标准力与测量范围10 - 2005年与意图的持续改进克罗地亚的计量基础设施的最后500千牛安装在LIMS。负载被液压获得和测量系统由与标称力100千牛和500千牛双置于参考换能器。初步的研究表明,相对测量不确定度将小于5E-04。进一步的调查电流。 LIMS已计划参加两个实验室之间的实验室间比对。使用100千牛和500个千牛标称值的两个商业力换能器500千牛 - 比对将在测量范围10来完成。参与实验室的最佳测量能力(CMC)从0,002percent变化到0,05percent。力换能器的校准将如DIN EN ISO 376比对描述将增加克罗地亚语计量的,从而提供进一步的合作研究水平使用实验室自己的校准过程中,相关的分类数据,即旋转效应,重复性,迟滞来进行在欧洲和全球国家实验室力关联。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号