首页> 外文会议>International Congress on Sound and Vibration >ANALYSIS OF DRIVING SOUND OF 1.5T MAGNETIC RESONANCE IMAGING (MRI) DEVICE IN CASE OF SLICE POSITIONING
【24h】

ANALYSIS OF DRIVING SOUND OF 1.5T MAGNETIC RESONANCE IMAGING (MRI) DEVICE IN CASE OF SLICE POSITIONING

机译:切片定位情况下1.5T磁共振成像(MRI)装置驱动声的分析

获取原文

摘要

The MRI examination is necessary and indispensable in the medical diagnosis. The generated operation sound is large at the image obtaining. The testee is pointing out dissatisfaction of it. The noise level is called 100dBA or more. This examination room was a strong magnetic field environment, and the noise level was not investigated enough. The reason is that measuring instrument breaks down when approaching the MRI device to measure it. In this report, the measurement result of the MRI driving sound using the non-magnetic microphone under the strong magnetic field is shown. MRI of GE Yokogawa medical system "SIGNA HORIZON LX 1.5T" has the gradient magnetic field of 20-70mT/m or less. The noise levels at some image obtaining were measured with this device. The maximum noise level of slice positioning was 102dBA. MRI of Philips Medical Systems "INTERA 1.5T" has the gradient magnetic field of 20-150mT/m or less. The noise levels at some image obtaining were measured with this device. The maximum noise level of slice positioning was 109dBA.
机译:MRI检查是必要的,在医学诊断中必不可少。在图像获取时生成的操作声音很大。 Testee指出对此的不满。噪声水平称为100dBA或更高。该检查室是一个强大的磁场环境,噪音水平没有足够的研究。原因是,当接近MRI设备测量它时,测量仪器会故障。在本报告中,示出了使用在强磁场下使用非磁性麦克风的MRI驱动声音的测量结果。 GE Yokogawa Medical System的MRI“Signa Horizo​​ n LX 1.5t”的梯度磁场为20-70mt / m或更小。使用该装置测量一些图像获得的噪声水平。切片定位的最大噪声水平为102dBa。飞利浦医疗系统的MRI“Intera 1.5t”的梯度磁场为20-150mt / m或更小。使用该装置测量一些图像获得的噪声水平。切片定位的最大噪声水平为109dBa。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号