7l Radiation dose to adult patients in LS spine X-ray examinations of health centres in one central hospital district in Finland

机译:7L辐射剂量对LS脊柱X射线检查的成人患者的芬兰中央医院医院医院X射线检查

获取原文

摘要

The present study on patient doses and examination techniques has shown that the doses exceed the national dose reference levels except in two health centres. The examination techniques don't meet the European quality criteria of focal spot size, total filtration, grid performance and nominal speed class of the image receptors in some health centres. The dose variations between the various health centres were found to be high, as were the dose ranges in a single health centre. The patient dose exceeded the national reference levels in only a few health centres, whereas in many health centres the examination techniques did not meet the European recommendations with regard to focal spot size and the speed index of the image receptor. Similar results with respect to the sensitivity of image receptors were reported earlier in health centres. The total dose measured in this study could be decreased by about 20% if more sensitive screens were used.
机译:目前对患者剂量和检查技术的研究表明,除了两个保健中心之外,剂量超出了国家剂量参考水平。检查技术不符合一些保健中心在图像受体的焦点尺寸,总滤波,网格性能和名义速度等级的欧洲质量标准。发现各种健康中心之间的剂量变化很高,因此单个保健中心的剂量范围是高的。患者剂量仅超过了几个卫生中心的国家参考水平,而在许多保健中心中,考试技术没有符合欧洲局部点尺寸和图像受体的速度指标的欧洲建议。在健康中心之前报道了相对于图像受体敏感性的结果。如果使用更敏感的屏幕,本研究中测量的总剂量可以降低约20%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号