Tolerance levels for quality assurance of electron density values generated for CT in radiotherapy treatment planning

机译:放射治疗计划中为CT生成的电子密度值的质量保证的公差等级

获取原文
获取原文并翻译 | 示例
           

摘要

Methods are described which relate the uncertainty in relative electron density derived from CT numbers to the uncertainty in treatment plan calculation for both photon and electron beams. These relationships are used to generate tolerance levels for electron density quality assurance measurements. These tolerance levels are dependent on treatment beam energy and tissue thickness, and are generally broader than current recommendations. The predicted treatment plan errors associated with these tolerance levels are shown to be consistent with calculations made using two commercial treatment planning systems. The tolerance levels are also shown to be practical by comparison against quality assurance measurements made on a conventional CT scanner and a treatment simulator CT system over a 12-month period. The results demonstrate that broader tolerances than are currently recommended can be justified in all situations except for electron beam treatments in which the therapeutic range falls within lung tissue.
机译:描述了将由CT数得出的相对电子密度的不确定性与光子束和电子束的治疗计划计算中的不确定性相关的方法。这些关系用于生成电子密度质量保证测量的公差等级。这些公差水平取决于治疗束能量和组织厚度,并且通常比当前推荐范围宽。与这些公差水平相关的预测治疗计划误差显示与使用两个商业治疗计划系统进行的计算一致。通过与常规CT扫描仪和治疗模拟器CT系统在12个月内进行的质量保证测量结果进行比较,也显示出公差水平是实用的。结果表明,除电子束治疗的治疗范围落在肺组织内外,在所有情况下都可以证明比当前建议的宽容忍度更大。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号