Effect of noise on T-wave alternans measurement in ambulatory ECGs using modified moving average versus spectral method.

机译:使用修正的移动平均对频谱方法,噪声对动态ECG中T波交替信号测量的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: The modified moving average (MMA) and spectral method (SM) are commonly used to measure T-wave alternans (TWA), but their accuracy has not been compared in ambulatory electrocardiograms (ECGs) where TWA signal-to-noise ratio is low. Our objective was to compare the effect of noise and signal nonstationarity on the accuracy of TWA measurement using MMA versus SM when applied to synthetic and ambulatory ECGs. METHODS: Periodic and nonperiodic noise were added to noiseless synthetic ECGs. Simulated TWA (0-20 microV) was added to synthetic ECGs and ambulatory ECG recordings. TWA was measured using SM and MMA, and the measurement error relative to added TWA was compared. An MMA ratio was used to discriminate TWA signal from noise. Signal nonstationarity was simulated by changing heart rate, TWA magnitude, and TWA phase. RESULTS: With no added TWA, MMA falsely measured TWA in synthetic and ambulatory ECGs, while false measurement was not seen with SM. An MMA ratio > 1.2 eliminated false TWA detection. In the presence of low TWA magnitude (<10 microV), TWA was overestimated by MMA and underestimated by SM in proportion to the noise level. In synthetic ECGs with periodic noise and 10-microV added TWA, MMA was less accurate than SM. The effects of simulated signal nonstationarity on the TWA magnitude measured with MMA versus SM were similar using a 64-beat analysis window. CONCLUSIONS: In the presence of noise, MMA falsely detects or overestimates simulated TWA in ambulatory ECG recordings. In this setting, the proposed MMA ratio improves the specificity of MMA.
机译:背景:修正的移动平均值(MMA)和频谱方法(SM)通常用于测量T波交替信号(TWA),但在TWA信噪比为的动态心电图(ECG)中,其准确性尚未得到比较。低。我们的目标是比较将噪声和信号非平稳性应用于合成和动态ECG时,使用MMA和SM进行的TWA测量准确性的影响。方法:将周期性和非周期性噪声添加到无噪声的合成心电图中。将模拟的TWA(0-20 microV)添加到合成ECG和动态ECG记录中。使用SM和MMA测量TWA,并比较相对于添加的TWA的测量误差。 MMA比用于将TWA信号与噪声区分开。通过改变心率,TWA幅度和TWA相位来模拟信号非平稳性。结果:在不添加TWA的情况下,MMA错误地测量了合成和动态ECG中的TWA,而SM未见错误的测量结果。 MMA比率> 1.2消除了错误的TWA检测。在低TWA幅度(<10 microV)的情况下,MMA与噪声水平成比例地高估了TWA,而SM低估了TWA。在具有周期性噪声和添加10-microV TWA的合成ECG中,MMA的准确性不如SM。使用64拍分析窗口,模拟信号非平稳性对MMA与SM测得的TWA幅度的影响相似。结论:在存在噪声的情况下,MMA错误地检测或高估了动态ECG记录中的模拟TWA。在这种情况下,建议的MMA比可提高MMA的特异性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号