A general Monte Carlo code for the simulation of energy dispersive X-ray fluorescence (ED-XRF) spectrometers is presented. As a significant extension to existing Monte Carlo codes, dedicated to ED-XRF spectrometers, the new code includes several unique features such as the simulation of M-lines and cascade effects. The detector response function was extended with fluorescence and Compton escape peaks, as well as with pulse pile-up. A full description of the underlying algorithms is given. The software was validated by means of comparison with experimental spectra of standard reference materials collected at the synchrotron XRF-microprobe installed at HASYLAB Beamline L, Hamburg, Germany.
机译:提出了用于模拟能量色散X射线荧光(ED-XRF)光谱仪的通用Monte Carlo代码。作为对专用于ED-XRF光谱仪的现有蒙特卡洛代码的重要扩展,新代码包括一些独特的功能,例如M线模拟和级联效应。检测器的响应功能随着荧光和康普顿逃逸峰以及脉冲堆积而扩展。给出了基础算法的完整描述。通过与在德国汉堡HASYLAB Beamline L上安装的Synchrontron XRF-microprobe收集的标准参考物质的实验光谱进行比较,对软件进行了验证。