首页> 外文期刊>Orthopedics >Thin-Walled Cross-Linked Acetabular Liners Need Not Exhibit Reduced Locking Strength
【24h】

Thin-Walled Cross-Linked Acetabular Liners Need Not Exhibit Reduced Locking Strength

机译:薄壁交联髋臼内衬不必表现出降低的锁定强度

获取原文
获取原文并翻译 | 示例
           

摘要

Use of larger diameter femoral heads has emerged as a promising strategy to reduce the risk of dislocation after total hip arthroplasty, but thinning the walls of crosslinked ultra-high-molecular-weight polyethylene (UHMWPE) acetabular liners to accommodate these larger heads may compromise the locking mechanism of the liner. The purpose of this study was to test the mechanical integrity of the locking mechanism in cross-linked and re-melted UHMWPE acetabular components with reduced wall thickness. The locking mechanism of cross-linked (100 kGy/re-melted) acetabular liners in sizes 50/28, 50/36, and 52/36 mm of 1 design was evaluated by lever-out tests and torsion tests. Torsion tests were performed at 2 angles to isolate the liner's locking tabs independent of the contribution of its central post. Lever-out testing demonstrated nominally reduced failure strength in 50/36-mm liners (13.3 N . m) compared with 50/28-mm liners (12.3 N . m; P=.0502), whereas the lever-out strength of 52/36-mm liners was 12.2 +/- 0.94 N . m. Failure torques were similar between 50/28-and 50/36-mm liners at 45 degrees and 90 degrees, but the failure torque of size 52/36-mm liners was significantly higher at each angle. The use of larger diameter femoral heads does not compromise the locking mechanism of thinned MicroSeal (Signal Medical Corp, Marysville, Michigan) acetabular liners. Use of a cross-linked UHMWPE acetabular liner, with a locking mechanism that is not compromised when the liner is thinned to a thickness of at least 2.86 mm, appears to be a biomechanically sound construct when articulated with large diameter femoral heads.
机译:使用大直径股骨头已成为减少全髋关节置换术后脱位风险的一种有前途的策略,但是要使交联的超高分子量聚乙烯(UHMWPE)髋臼内衬的壁变薄以适应这些较大的股骨头可能会损害衬套的锁定机构。这项研究的目的是测试壁厚减小的交联和重熔UHMWPE髋臼组件中锁定机构的机械完整性。通过杠杆试验和扭转试验评估了1种设计尺寸为50 / 28、50 / 36和52/36 mm的交联(100 kGy /重熔)髋臼内衬的锁定机制。扭转测试以2个角度进行,以隔离衬管的锁定凸耳,而不受其中央支柱的影响。杠杆测试表明,与50/28毫米衬套(12.3 N.m; P = .0502)相比,50/36毫米衬套(13.3 N.m)名义上的破坏强度降低了,而杠杆强度为52 / 36毫米内衬为12.2 +/- 0.94N。米在45度和90度时,50 / 28-mm和50 / 36-mm衬套的破坏扭矩相似,但是52 / 36-mm尺寸的衬套的破坏扭矩在每个角度均明显更高。较大直径的股骨头的使用不会损害较薄的MicroSeal(Signal Medical Corp,密歇根州马里斯维尔)髋臼内衬的锁定机制。交联的UHMWPE髋臼内衬的厚度薄到至少2.86 mm时,锁定机制不会受到损害,当与大直径股骨头相连时,似乎是一种生物力学上可靠的构造。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号