首页> 外文期刊>Ophthalmic Surgery and Lasers >Bacterial contamination of the pressure receiver of a vitrectomy machine.
【24h】

Bacterial contamination of the pressure receiver of a vitrectomy machine.

机译:玻璃体切割机压力接收器的细菌污染。

获取原文
获取原文并翻译 | 示例
           

摘要

On a routine sterility check, the authors found bacteria in their vitrectomy machine. Serratia marcescens was detected in the pressure receiver of a roller pump vitrectomy machine. The authors improved the machine by installing a millipore filter in front of the receiver that prevents bacteria from entering it without interfering with the pressure measurements. The origin of Serratia was probably a patient with endophthalmitis who was operated on long before the routine check. Although this appears to be the first report of a bacterial contamination of a vitrectomy machine, the results of testing show that this situation is indeed possible. The authors strongly recommend routine checks of the equipment and suggest mounting an antibacterial filter.
机译:在常规的无菌检查中,作者在其玻璃体切割机中发现了细菌。在滚子泵玻璃体切割机的压力接收器中检测到粘质沙雷氏菌。作者通过在接收器的前面安装微孔过滤器来改进机器,以防止细菌进入机器而不会干扰压力测量。沙雷氏菌的起源可能是眼内炎患者,在常规检查之前很久就接受了手术。尽管这似乎是玻璃体切割机细菌污染的首次报道,但测试结果表明,这种情况确实是可能的。作者强烈建议对设备进行例行检查,并建议安装抗菌过滤器。