This article presents results of measurements for the calibration of radiation protection area monitoring instruments carried out during the period 2006-2008 at Secondary Standard Dosimetry Laboratory of Sudan. The work performed included quality assurance measurements, measurements for the dosimeter calibrations and uncertainty analysis. Calibrations were performed using 137Cs gamma ray sources produced by OB 85 and OB 34/1 gamma calibrators producing air kerma rate that ranged from 10 μGy/h to 50 mGy/h. Area monitoring instruments were calibrated in terms of ambient dose equivalent, H (*)(10) derived using air kerma to ambient dose equivalent conversion coefficients. Results are presented for 78 area monitoring instruments representing most commonly used types in Sudan. Radioactive check source measurements for the reference chamber showed deviation within 1% limit. The accuracy in the beam output measurements was within 5% internationally considered as acceptable. The results highlighted the importance of radiation protection calibrations. Regulations are further need to ensure safety aspect really meet the required international standards.
机译:本文介绍了在苏丹二级标准剂量测定实验室进行的2006-2008年期间进行的辐射防护区域监测仪器校准的测量结果。进行的工作包括质量保证测量,剂量计校准测量和不确定性分析。使用OB 85和OB 34/1伽马校准器产生的137Cs伽马射线源进行校准,产生的空气比释动能范围为10μGy/ h至50 mGy / h。使用空气比释动能到环境剂量当量的转换系数,根据环境剂量当量H(*)(10)对面积监测仪器进行了校准。列出了代表苏丹最常用类型的78种区域监测仪器的结果。参考室的放射性检查源测量结果显示偏差在1%的限制内。光束输出测量的精度在国际公认的5%以内。结果突出了辐射防护校准的重要性。还需要制定法规以确保安全方面真正满足要求的国际标准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号