摘要

We present an overview of accelerator mass spectrometry (AMS) radiocarbon sample preparation and measurements, describing the technical upgrades that now allow us to routinely obtain 0.2-0.3% precision for l-mg carbon samples. A precision of similar to 1 % on samples with 100 mu g of carbon can also be achieved. We have also developed graphitization techniques and AMS procedures for ultra-small samples (down to 0.002 mg of carbon). Detailed time series are presented for large and small aliquots of standards such as NIST OX-I and OX-II; FIRI-C and -D; IAEA-C6, -C7 and -C8; and C-14-free samples.
机译:我们对加速器质谱(AMS)放射性碳样品的制备和测量进行了概述,描述了技术升级,现在我们可以常规地获得1-mg碳样品的0.2-0.3%的精度。对于含100μg碳的样品,也可以达到接近1%的精度。我们还为超小型样品(低至0.002 mg碳)开发了石墨化技术和AMS程序。给出了详细的时间序列,其中包括NIST OX-I和OX-II等标准品的大小。 FIRI-C和-D; IAEA-C6,-C7和-C8;和不含C-14的样品。