首页> 外文期刊>Retina >The effect of transducer bandwidth on ultrasonic image characteristics.
【24h】

The effect of transducer bandwidth on ultrasonic image characteristics.

机译:换能器带宽对超声图像特征的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: This study was conducted to determine the effect of transducer bandwidth on the characteristics of ophthalmic ultrasound images. METHODS: B-scan images produced using two transducers, one with a narrow bandwidth and the other with a broad bandwidth, both having nominal center frequencies of 10 MHz, were evaluated. Comparative scans were made of a tissue-mimicking phantom, an intraocular tumor, and a vitreous hemorrhage. RESULTS: Results showed that broadband transducers gave improved resolution and finer speckle texture, but had lower sensitivity. Broadband transducers were most suitable for situations in which resolution was more important than sensitivity, such as imaging of tumors. The greater sensitivity of the narrowband probe made it most useful for evaluation of vitreous complications, such as hemorrhage and membranes. CONCLUSION: In addition to transducer frequency, bandwidth should be taken into account when choosing a transducer for a specific ophthalmic imaging application. Both broad- and narrowband transducers have relative advantages in particular applications.
机译:目的:进行这项研究以确定换能器带宽对眼科超声图像特征的影响。方法:评估使用两个换能器产生的B扫描图像,一个换能器的带宽很窄,另一个换能器的带宽很宽,都具有10 MHz的标称中心频率。比较扫描组织模仿体模,眼内肿瘤和玻璃体出血。结果:结果表明,宽带换能器具有更高的分辨率和更细的斑点纹理,但灵敏度较低。宽带换能器最适合分辨率比灵敏度更重要的情况,例如肿瘤成像。窄带探头的灵敏度更高,使其最适合用于评估玻璃体并发症,例如出血和膜。结论:在为特定眼科成像应用选择换能器时,除了换能器频率外,还应考虑带宽。宽带和窄带换能器在特定应用中均具有相对优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号