The 6 circular grating eccentricity errors model attempts to improve the measurement accuracy of an articulated arm coordinate measuring machine (AACMM) without increasing the corresponding hardware cost. We analyzed the AACMM's circular grating eccentricity and obtained the 6 joints' circular grating eccentricity error model parameters by conducting circular grating eccentricity error experiments. We completed the calibration operations for the measurement models by using home-made standard bar components. Our results show that the measurement errors from the AACMM's measurement model without and with circular grating eccentricity errors are 0.0834 mm and 0.0462 mm, respectively. Significantly, we determined that measurement accuracy increased by about 44.6% when the circular grating eccentricity errors were corrected. This study is significant because it promotes wider applications of AACMMs both in theory and in practice.
机译:6个圆光栅偏心误差模型试图在不增加相应硬件成本的情况下提高关节臂坐标测量机(AACMM)的测量精度。通过进行圆光栅偏心误差实验,分析了AACMM的圆光栅偏心率,得到了6个关节的圆光栅偏心误差模型参数。我们使用自制的标准条形组件完成了测量模型的校准操作。我们的结果表明,AACMM测量模型在没有和有圆形光栅偏心误差的情况下的测量误差分别为0.0834 mm和0.0462 mm。重要的是,当校正圆光栅偏心误差时,我们确定测量精度提高了约44.6%。这项研究意义重大,因为它在理论和实践上都促进了AACMM的广泛应用。