Use of the U-clip for microvascular anastomosis.

机译:使用U型夹进行微血管吻合。

获取原文
获取原文并翻译 | 示例
           

摘要

Microvascular anastomosis is a demanding skill requiring technical excellence and a thorough knowledge of anatomy and physiology. Every suture placed in a microvascular anastomosis should be considered critical as each has the potential to compromise the delicate reconstruction. As such, any device that can facilitate microvascular suture placement deserves thorough evaluation. The U-clip (Coalescent Surgical, Sunnyvale, CA) is such a device in that it eliminates the often time consuming process of tying knots. We evaluated use of the U-clip in microvascular anastomosis of a 1.5 mm artery. We found the U-clip to offer some advantages including ease of use, traditional feel of directed suture placement (as compared to couplers), and elimination of time needed for knot tying. Its shortcomings include size (in diameter, the "pop-off" section of the device appears larger than standard 8-0 suture), the significant force required to "pop-off" the device and difficulty removing the device.
机译:微血管吻合术是一项要求很高的技能,需要技术卓越以及对解剖学和生理学的全面了解。放置在微血管吻合中的每条缝合线都应视为关键,因为每条缝合线都有可能损害微妙的重建效果。这样,任何能够促进微血管缝合线放置的装置都应进行全面评估。 U形夹(Coalescent Surgical,Sunnyvale,CA)就是这样一种装置,因为它消除了通常很费时的打结过程。我们评估了U型夹在1.5毫米动脉微血管吻合中的使用。我们发现U形夹具有一些优势,包括易用性,定向缝线放置的传统感觉(与耦合器相比)以及消除打结所需的时间。它的缺点包括尺寸(直径,设备的“弹出”部分看上去比标准的8-0缝线大),“弹出”设备所需的很大的力以及难以移除设备。