首页> 外文期刊>Microchemical Journal: Devoted to the Application of Microtechniques in all Branches of Science >Determination of copper in medicinal plants used as dietary supplements by atomic absorption spectrometry with direct flame solid analysis
【24h】

Determination of copper in medicinal plants used as dietary supplements by atomic absorption spectrometry with direct flame solid analysis

机译:原子吸收光谱法直接火焰固体分析法测定膳食补充剂药用植物中的铜

获取原文
获取原文并翻译 | 示例
           

摘要

An alternative device for the direct solid analysis (DSA) for copper determination by flame atomic absorption spectrometry (FAAS) is proposed. Copper was directly determined in commercial medicinal plants used as dietary supplements. The determination of copper in solid samples by DSA-FAAS was made by using a conventional air-acetylene flame. Between 0.05 and 1.5 mg of each test, sample was weighed directly into a small polyethylene vial connected to the device used for solid introduction into the flame. Test samples were introduced into the flame as a dry aerosol using a T-quartz cell set between the burner and the optical path. The T-quartz cell has a slit in the superior part by which the solid aerosol passes to the flame. A transient signal, evaluated as integrated absorbance, is produced and it is totally integrated in 2 s. Background signals always presented absorbance values less than 0.1. It was found a characteristic mass of 0.8 ng Cu and absolute limit of detection of 1.2 ng (3 s), or 1.2 mug g(-1) if a sample mass of 1 mg was used. Optimized conditions for air flow rate, flame stoichiometry, and so on were established as well. No excessive grinding of the samples was needed and samples with particle of size less than 80 Pm were used throughout. No statistical difference between the results from the proposed system and those obtained by sample digestion and determination by conventional FAAS was observed. With the proposed procedure, more than 50 test samples can be analyzed in 1 h and it can be easily adapted to conventional spectrometers for FAAS. (C) 2004 Elsevier B.V. All rights reserved.
机译:提出了一种用于通过火焰原子吸收光谱法(FAAS)测定铜的直接固体分析(DSA)的替代设备。铜是在用作膳食补充剂的商业药用植物中直接测定的。使用常规的空气乙炔火焰,通过DSA-FAAS测定固体样品中的铜。在每次测试的0.05至1.5 mg之间,将样品直接称量到一个小的聚乙烯小瓶中,该小瓶与用于将固体引入火焰的装置相连。使用设置在燃烧器和光路之间的T石英池将测试样品作为干燥的气溶胶引入火焰中。 T石英池的上部有一个缝隙,固体气溶胶通过该缝隙进入火焰。产生一个瞬态信号,被评估为积分吸光度,并且在2 s内将其完全积分。背景信号的吸光度值始终小于0.1。如果使用1 mg的样品质量,发现特征质量为0.8 ng Cu,绝对检测极限为1.2 ng(3 s),即1.2杯子g(-1)。还建立了空气流速,火焰化学计量等的最佳条件。无需过度研磨样品,并且始终使用粒径小于80 Pm的样品。在提议的系统的结果与通过样品消解和常规FAAS测定获得的结果之间没有观察到统计学差异。通过提出的程序,可以在1小时内分析50多个测试样品,并且可以轻松地将其应用于FAAS的常规光谱仪。 (C)2004 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号