BACKGROUND: This study compared the fluid leakage in the new 'tapered' shaped against the classic 'cylindrical' shaped tracheal tube cuffs when placed in different sized tracheas. METHODS: The 7.5 mm internal diameter (ID) tracheal tube cuffs-Tapered Seal Guard (TSG), Standard Seal Guard (SSG), Hi-Lo, Microcuff, Ruesch, and Portex Profile-were compared in an in vitro apparatus. Vertical artificial tracheas with 16, 20, and 22 mm ID were intubated, 5 ml clear water was applied above the unlubricated tube cuffs, and fluid leakage was measured up to 60 min. Data of tapered vs non-tapered tube cuffs (16 observations) were compared for each tracheal diameter using the Mann-Whitney test. RESULTS: Median (range) fluid leakage (ml) at 60 min was 2.14 (0.05-4.88), 1.14 (0.00-4.84), and 0.13 (0.00-1.32), respectively, for 16, 20, and 22 mm tracheas in the TSG tube studies when compared with 4.58 (0.44-4.88), 2.21 (0.00-4.81), and 0.00 (0.00-4.81) in the SSG tube and 4.54 (1.54-4.82), 0.90 (0.00-4.49), and 4.85 (4.40-4.99) in the Microcuff tube studies. Leakage in all polyvinylchloride (PVC) tube cuffs was almost complete (5 ml) within 5 min (P<0.001). CONCLUSIONS: The tapered PU tube cuff was as effective as the cylindrical PU cuffs in smaller tracheal diameters and was more efficient than the cylindrical Microcuff PU tube cuff in larger tracheal diameter in preventing subglottic fluid leakage across the tube cuff tested in this in vitro study. PVC tube cuffs leaked much more and faster than PU cuffs.
机译:背景:这项研究比较了当放置在不同大小的气管中时,新的“锥形”形与经典的“圆柱形”形气管插管的漏液情况。方法:在体外装置中比较了7.5 mm内径(ID)气管导管袖带-锥形密封护罩(TSG),标准密封护罩(SSG),高-低,Microcuff,Ruesch和Portex Profile。插入具有16、20和22毫米内径的垂直人工气管,在未润滑的管套上方加5毫升清水,并测量长达60分钟的液体渗漏。使用Mann-Whitney检验比较每个气管直径的锥形和非锥形管套囊的数据(16个观察结果)。结果:对于16 mm,20 mm和22 mm的气管,在60分钟时的中位(范围)渗漏量(ml)分别为2.14(0.05-4.88),1.14(0.00-4.84)和0.13(0.00-1.32)。与SSG管中的4.58(0.44-4.88),2.21(0.00-4.81)和0.00(0.00-4.81)以及4.54(1.54-4.82),0.90(0.00-4.49)和4.85(4.40)相比,TSG管研究-4.99)在Microcuff管研究中。在5分钟内,所有聚氯乙烯(PVC)管套的泄漏几乎完全(5 ml)(P <0.001)。结论:在该体外研究中,锥形PU管袖套与较小气管直径的圆柱PU袖套一样有效,并且比较大直径气管直径的圆柱形Microcuff PU管套更有效,可以防止声门下液通过管套漏出。 PVC管袖口的泄漏比PU管袖的泄漏要快得多。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号