Kilovoltage cone-beam CT imaging dose during breast radiotherapy: A dose comparison between a left and right breast setup

机译:乳房放射治疗中的千伏锥束CT成像剂量:左右乳房设置之间的剂量比较

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to investigate the delivered dose from a kilovoltage cone-beam computed tomography (kV-CBCT) acquired in breast treatment position for a left and right breast setup. The dose was measured with thermoluminescent dosimeters positioned within a female anthropomorphic phantom at organ locations. Imaging was performed on an Elekta Synergy XVI system with the phantom setup on a breast board. The image protocol involved 120. kVp, 140. mAs, and a 270° arc rotation clockwise 0° to 270° for the left breast setup and 270° to 180° for the right breast setup (maximum arc rotations possible). The dose delivered to the left breast, right breast, and heart was 5.1. mGy, 3.9. mGy, and 4.0. mGy for the left breast setup kV-CBCT, and 6.4. mGy, 6.0. mGy, and 4.8. mGy for the right breast setup kV-CBCT, respectively. The rotation arc of the kV-CBCT influenced the dose delivered, with the right breast setup kV-CBCT found to deliver a dose of up to 4. mGy or 105% higher to the treated breast's surface in comparison with the left breast setup. This is attributed to the kV-CBCT source being more proximal to the anterior of the phantom for a right breast setup, whereas the source is more proximal to the posterior of the patient for a left-side scan.
机译:这项研究的目的是研究在左右乳房设置中在乳房治疗位置获得的千伏锥束计算机断层扫描(kV-CBCT)的输送剂量。用位于器官拟人化拟人体模内的热辐射剂量计测量剂量。成像是在Elekta Synergy XVI系统上进行的,幻像设置在胸板上。图像协议涉及120. kVp,140。mAs,左乳房设置为270°顺时针旋转0°至270°,右乳房设置为270°至180°(可能最大旋转)。输送到左乳房,右乳房和心脏的剂量为5.1。 mGy,3.9。 mGy和4.0。左侧乳房设置kV-CBCT的mGy和6.4。 mGy,6.0。 mGy和4.8。分别针对右乳房设置kV-CBCT的mGy。 kV-CBCT的旋转弧影响输送的剂量,与左侧的乳房设置相比,右侧的乳房设置kV-CBCT发现向治疗的乳房表面输送的剂量最多高4 mGy或105%。对于右乳房设置,这归因于kV-CBCT源更靠近体模的前部,而对于左侧扫描,该源更靠近患者的后部。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号