Setup errors in patients treated with intensity-modulated whole pelvic radiation therapy for gynecological malignancies.

机译:妇科恶性肿瘤接受强度调制全盆腔放射治疗的患者的设置错误。

获取原文
获取原文并翻译 | 示例
           

摘要

Intensity-modulated whole pelvic radiation therapy (IM-WPRT) has decreased the incidence of gastrointestinal complications by reducing the volume of normal tissue irradiated in gynecologic patients. However, IM-WPRT plans result in steep dose gradients around the target volume, and thus accurate patient setup is essential. To quantify the accuracy of our patient positioning, we examined the weekly portal films of 46 women treated with IM-WPRT at our institution. All patients were positioned using a customized immobilization device that was indexed to the treatment table. Setup errors were evaluated by comparing portal images to simulation images using an algorithm that registers user-defined open curve segments drawn on both sets of film. The setup errors, which were separated into systematic and random components, ranged from 1.9 to 3.7 mm for the translations and 1.3 degrees to 4.4 degrees for the 2 in-plane translations. The systematic errors were all less than the respective random errors, with thelargest error in the anterior/posterior direction. In addition, there was no correlation between the magnitude of these errors and patient-specific factors (age, weight, height). In the future, we will investigate the effect of these setup errors on the delivered dose distribution.
机译:调强全盆腔放射治疗(IM-WPRT)通过减少妇科患者的正常组织照射量来降低胃肠道并发症的发生率。但是,IM-WPRT计划会导致目标体积附近出现陡峭的剂量梯度,因此准确的患者设置至关重要。为了量化患者定位的准确性,我们在我们的机构检查了46例接受IM-WPRT治疗的女性的每周门诊影片。使用定制的固定装置对所有患者进行定位,该固定装置被索引到治疗台上。通过使用将记录在两组胶卷上绘制的用户定义的开放曲线段的算法进行比较,通过将门户图像与模拟图像进行比较来评估设置错误。设置误差分为系统性和随机性两个部分,平移范围为1.9毫米至3.7毫米,两次平面内平移的范围为1.3度至4.4度。系统误差均小于各自的随机误差,前/后方向的误差最大。此外,这些误差的大小与患者特定因素(年龄,体重,身高)之间没有关联。将来,我们将研究这些设置误差对所输送剂量分布的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号