首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >The OxyArm - a new minimal contact oxygen delivery system for mouth or nose breathing.
【24h】

The OxyArm - a new minimal contact oxygen delivery system for mouth or nose breathing.

机译:OxyArm-一种用于嘴或鼻子呼吸的新型最小接触氧气输送系统。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To describe the development and performance of a new minimal contact oxygen (O(2)) delivery system for both nasal and oral breathing, with capnographic capabilities. METHODS: The development and design challenges of the OxyArm (OA) prototype are described. The innovative design utilizes a headset with a semi-rigid boom and an O(2) diffuser. The OA was compared to the Venturi mask in eight healthy adults after informed consent. Inspired O(2) fractions were measured in the hypopharynx using continuous gas sampling at low to high O(2) flow rates. Mean data were compared using two-tailed paired t tests with significance set at 0.05. RESULTS: The measured inspired O(2) concentration was higher in the OA at 2 (26.3 +/- 2.5 vs 23.3 +/- 0.5, P <0.01) and 6 L x min(-1) (33.5 +/- 3.3 vs 28.8 +/- 1.2, P <0.01) flow rates. At 12 L x min(-1), the O(2) concentration was less in the OA (39.2 +/- 6.3 vs 46.0 +/- 2.7, P <0.02). All subjects found both systems comfortable for the short duration of the study. CONCLUSIONS: The OA delivered predictable concentrations of O(2) over low to medium flow rates. This system is comfortable, easy to use, non-obtrusive, odorless, and latex-free. The ability to monitor capnography makes this device ideal for monitored anesthesia care or in other settings (intensive care) where monitoring of respiration is warranted. This device does not contact the face and thus may be ideal for pediatric patients and those on long-term home O(2) therapy. Further clinical trials in these areas are warranted.
机译:目的:描述具有二氧化碳分析功能的用于鼻呼吸和口呼吸的新型最小接触氧气(O(2))输送系统的开发和性能。方法:描述了OxyArm(OA)原型的开发和设计挑战。创新的设计采用了具有半刚性吊杆和O(2)扩散器的头戴式耳机。在知情同意后,将八名健康成人的OA与文丘里面膜进行了比较。使用低至高O(2)流速的连续气体采样在下咽中测量被激发的O(2)馏分。使用两尾配对t检验比较平均数据,显着性设为0.05。结果:在2(26.3 +/- 2.5 vs 23.3 +/- 0.5,P <0.01)和6 L x min(-1)(33.5 +/- 3.3 vs 28.8 +/- 1.2,P <0.01)流速。在12 L x min(-1)时,OA中的O(2)浓度较低(39.2 +/- 6.3对46.0 +/- 2.7,P <0.02)。所有受试者都认为这两个系统在短期研究中均很舒适。结论:OA在低至中等流速下提供了可预测的O(2)浓度。该系统舒适,易用,无干扰,无味且不含乳胶。二氧化碳分析仪的监视功能使其成为理想的监护麻醉护理或需要监测呼吸的其他环境(重症监护)。该设备不会接触面部,因此对于小儿患者和长期在家进行O(2)治疗的患者可能是理想的选择。有必要在这些领域进行进一步的临床试验。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号