Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia.

机译:LASIK术后近视眼中使用Pentacam和超声测厚仪测量中央角膜厚度的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To compare central corneal thickness (CCT) measurements obtained using the Pentacam Scheimpflug system with those obtained using DGH ultrasound pachymetry (UP) in post-laser in situ keratomileusis (LASIK) eyes for myopia. METHODS: In a prospective study, measurement agreement was assessed in 143 eyes of 72 post-LASIK patients using both the Pentacam and UP at the Eye Consultants Center in Riyadh, Saudi Arabia. RESULTS: The mean CCT was 522 +/- 42.2 microm with the Oculus Pentacam and 516.2 +/- 40.6 microm with UP. The Bland-Altman plot showed that the mean +/- SD for the differences between the 2 devices was 5.8 +/- 13.6 microm, with 95% confidence interval limits ranging from -20.9 microm to 32.6 microm. A test of statistical significance indicated that the mean differences of 5.8 +/- 13.6 microm differed significantly from zero (p<0.001; Wilcoxon signed-rank test), thus indicating that the Pentacam measurements tended to overestimate CCT compared with UP. Analysis of regression showed a high correlation between the values obtained with both devices (r=0.947, p<0.001). CONCLUSIONS: In post-LASIK myopic eyes, although a high correlation has been shown between Pentacam and UP measurements, Pentacam tends to overestimate CCT compared to UP. Pentacam probably cannot be used interchangeably with UP in post-LASIK eyes for myopia.
机译:目的:比较使用Pentacam Scheimpflug系统获得的角膜中央厚度(CCT)测量值和使用DGH超声测厚法(UP)获得的近视激光后原位角膜磨镶术(LASIK)眼睛测量值。方法:在一项前瞻性研究中,在沙特阿拉伯利雅得的眼科咨询中心使用Pentacam和UP对72例LASIK术后患者的143只眼的测量协议进行了评估。结果:Oculus Pentacam的平均CCT为522 +/- 42.2微米,UP的平均CCT为516.2 +/- 40.6微米。 Bland-Altman图显示,两个设备之间的差异的平均值+/- SD为5.8 +/- 13.6微米,95%的置信区间范围为-20.9微米至32.6微米。统计学显着性检验表明,平均差异5.8 +/- 13.6微米与零显着不同(p <0.001; Wilcoxon符号秩检验),因此表明Pentacam测量与UP相比倾向于高估CCT。回归分析显示,两种设备获得的值之间具有高度相关性(r = 0.947,p <0.001)。结论:在LASIK术后近视眼中,尽管Pentacam和UP测量之间显示出高度相关性,但与UP相比,Pentacam倾向于高估CCT。 Pentacam可能无法在LASIK后近视眼中与UP互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号