OBJECTIVES: A feasibility study. METHODS: Eight pigs (all females; mean weight: 29 kg) underwent a conventional transperitoneal aortic approach with implantation of an aorto-prosthetic end-to-side anastomosis using a Clampless device and deployment of a 5-mm polytetrafluoroethylene (PTFE) graft. After proximal ligature, a conventional end-to-end anastomosis was then performed between the graft and the left iliac artery. RESULTS: The first pig died during the procedure due to graft misplacement. The seven other procedures were successful with a mean operative and anastomosis time of 101 min (range: 81-115 min) and 3.35 min (range: 2.25-4.25 min), respectively; mean blood loss was 152 ml (range: 30-235 ml). Another pig with a patent graft died at day 4 as a result of a severe unrelated pneumonopathy. The angiogram performed during the procedure and before sacrifice, at 2 (n=2), 4 (n=2) and 6 weeks (n=2), showed no graft stenosis or thrombosis. Microscopic examination revealed a tissue covering the intraluminal stent, which evolved over time, with no visible endothelial proliferation or inflammation. CONCLUSION: An aorto-prosthetic anastomosis can be performed safely and efficiently with our new clampless and sutureless device. The next step will be a laparoscopic Clampless implantation.
机译:目的:可行性研究。方法:八只猪(全部为雌性;平均体重:29千克)采用常规的经腹膜主动脉入路,使用无钳装置植入主动脉假体端对侧吻合术,并部署5毫米聚四氟乙烯(PTFE)移植物。在近端结扎之后,然后在移植物和左动脉之间进行常规的端到端吻合。结果:第一头猪在手术过程中因移植错位而死亡。其他七个手术均成功完成,平均手术时间和吻合时间分别为101分钟(范围:81-115分钟)和3.35分钟(范围:2.25-4.25分钟);平均失血量为152毫升(范围:30-235毫升)。另一只移植有专利的猪在第4天死于严重的无关肺病。在手术过程中和处死前的第2(n = 2),4(n = 2)和6周(n = 2)进行的血管造影显示无移植物狭窄或血栓形成。显微镜检查发现覆盖腔内支架的组织随着时间的流逝而发展,没有可见的内皮增生或炎症。结论:使用我们的新型无钳无缝合装置可以安全有效地进行主动脉假体吻合。下一步将是腹腔镜无钳植入术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号