首页> 外文期刊>European journal of clinical microbiology and infectious diseases: Official publication of the European Society of Clinical Microbiology >Comparison of the Micronaut Merlin automated broth microtiter system with the standard agar dilution method for antimicrobial susceptibility testing of mucoid and nonmucoid Pseudomonas aeruginosa isolates from cystic fibrosis patients.
【24h】

Comparison of the Micronaut Merlin automated broth microtiter system with the standard agar dilution method for antimicrobial susceptibility testing of mucoid and nonmucoid Pseudomonas aeruginosa isolates from cystic fibrosis patients.

机译:比较Micronaut Merlin自动化肉汤微量滴定系统与标准琼脂稀释方法对囊性纤维化患者粘液样和非粘液性铜绿假单胞菌菌株的药敏试验。

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this study was to compare a commercially available automated broth microdilution system (Merlin; Micronaut, Germany) with the standard agar dilution method for susceptibility testing of pulmonary isolates from cystic fibrosis patients. Accurate susceptibility testing of bacterial isolates from cystic fibrosis patients is known to pose problems. Although commercially available automated test systems could facilitate susceptibility testing of such isolates in routine diagnostics, these systems have not been recommended thus far. However, a pilot study recently indicated that the Merlin system, which is based on an endpoint measurement rather than on growth curves, might be applicable in the susceptibility testing of isolates from cystic fibrosis patients. In the present study, the Merlin system was further evaluated using an extended panel of nonmucoid and mucoid Pseudomonas aeruginosa isolates. The results showed that the MICs obtained with the Merlin system tended to be lower than those obtained with the agar dilution method, a finding that became increasingly apparent when mucoid Pseudomonas aeruginosa strains were tested. The correlation coefficients (r values) of the MIC results for all strains tested were between 0.6 and 0.8 for five of the seven antimicrobial agents, with r values exceeding 0.8 for only meropenem and ciprofloxacin. However, since the overall rate of serious discrepancies was within an acceptable range, the Merlin system appears to be applicable for routine clinical use in susceptibility testing of P. aeruginosa isolates from cystic fibrosis patients.
机译:这项研究的目的是比较市售的自动化肉汤微量稀释系统(Merlin; Micronaut,德国)与标准琼脂稀释方法对囊性纤维化患者肺分离物的药敏试验。已知对来自囊性纤维化患者的细菌分离物进行准确的药敏测试会带来问题。尽管市售的自动化测试系统可以促进常规诊断中此类分离物的药敏性测试,但到目前为止,尚未推荐使用这些系统。然而,最近的一项初步研究表明,基于终点测量而非生长曲线的Merlin系统可能适用于囊性纤维化患者分离株的药敏试验。在本研究中,使用扩展的非粘液性和粘液性铜绿假单胞菌分离物评估了Merlin系统。结果表明,用Merlin系统获得的MIC倾向于低于通过琼脂稀释法获得的MIC,这一发现在测试黏液状铜绿假单胞菌菌株时变得越来越明显。对于七种抗菌药物中的五种,所有测试菌株的MIC结果的相关系数(r值)在0.6到0.8之间,而对于美罗培南和环丙沙星,r值超过0.8。但是,由于严重差异的总体发生率在可接受的范围内,因此,Merlin系统似乎适用于常规临床应用,用于对来自囊性纤维化患者的铜绿假单胞菌分离物进行药敏试验。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号