A simple in vitro method for the evaluation of an ultrasonic nebulizer for drug delivery to intubated, ventilated patients and the effect of nebulizer and ventilator settings on the uptake of fluid from the nebulizer chamber.

机译:一种简单的体外方法,用于评估超声雾化器向插管通气患者的药物输送情况,以及雾化器和呼吸机设置对从雾化室吸收液体的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

Little is known about the performance of ultrasonic nebulizers during different ventilation patterns when these nebulizers are used to deliver drugs to intubated, ventilated patients. A method that enables the performance of an ultrasonic nebulizer to be evaluated is described. We used an in vitro model to examine the performance of the DeVilbiss Ultra-Neb 2000 ultrasonic nebulizer under positive pressure ventilation. Performance was measured at different rates of nebulization and under changing conditions of positive end-expiratory pressure (PEEP), inspiratory flow rate, inspiratory time and minute volume. The volume of saline nebulized was unchanged by variations in positive end expiratory pressure from 0 to 5 cm to 10 cm H2O, in minute ventilation and in inspiratory flow rate. An increase in the inspiratory time resulted in an increase in the volume of saline nebulized and this volume was greater as the power setting of the nebulizer was increased. We conclude that ultrasonic nebulizers may be affected by different patterns of ventilation and that this simple in vitro assessment of nebulizer function in an intensive care setting may be of value prior to nebulizer use.
机译:当这些超声雾化器用于向插管通气的患者输送药物时,对于不同的通气模式,超声雾化器的性能知之甚少。描述了一种能够评估超声雾化器性能的方法。我们使用体外模型来检查DeVilbiss Ultra-Neb 2000超声雾化器在正压通风下的性能。在不同的雾化速率下以及在呼气末正压(PEEP),吸气流速,吸气时间和分钟体积变化的条件下测量性能。雾化盐水的体积在0至5 cm至10 cm H2O的呼气末正压变化,分钟通气量和吸气流速方面没有变化。吸气时间的增加导致雾化盐水的体积增加,并且随着雾化器功率设置的增加,该体积更大。我们得出的结论是,超声雾化器可能会受到不同通气方式的影响,并且在重症监护环境中对雾化器功能进行的这种简单的体外评估可能在使用雾化器之前具有一定的价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号