首页> 外文期刊>British Journal of Radiology >Investigation of dose homogeneity in paediatric anthropomorphic phantoms for a simple total body irradiation technique.
【24h】

Investigation of dose homogeneity in paediatric anthropomorphic phantoms for a simple total body irradiation technique.

机译:儿科拟人幻影中剂量均匀性的简单全身照射技术研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The technique for treating total body irradiation patients used at the centre involves no compensation for the inhomogeneity of patient shape. Dose is prescribed to the lung, and monitor units are derived from standard data depending on the external dimensions of the patient at nipple level. Dose measurements were made during standard treatments on three paediatric anthropomorphic phantoms representing children of 5, 10 and 15 years of age. The results confirmed that the measured dose to the lung was within 4% of the prescribed dose, and dose homogeneity was within +/- 5%, excluding the neck, where the higher measured doses were still within tissue tolerance.
机译:在中心使用的治疗全身照射患者的技术不补偿患者形状的不均匀性。规定了肺部的剂量,并且根据患者乳头水平的外部尺寸,从标准数据中得出监测单位。在标准治疗期间对代表5岁,10岁和15岁儿童的三个小儿拟人化体模进行剂量测量。结果证实,对肺部的测量剂量在规定剂量的4%之内,并且剂量均一性在+/- 5%之内(不包括颈部),其中较高的测量剂量仍在组织耐受范围内。