首页> 外文期刊>International journal of laboratory hematology >Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusion.
【24h】

Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusion.

机译:使用三种自动血液分析仪与免疫血小板参考方法进行血小板计数评估,并评估可能的血小板输注不足。

获取原文
获取原文并翻译 | 示例
           

摘要

The accuracy of three automated haematology analysers [Sysmex XE-2100 (both optical and impedance mode), Bayer Advia 120, and Beckman Coulter LH-750] was compared with the immunoplatelet reference method for platelet measurement. A total of 165 blood specimens were obtained from patients and platelet counts were determined using the four-automated haematology analyser methods and the immunoplatelet reference method. The coefficients of determination (R(2)) between the automated haematology analyser methods and the immunoplatelet reference method for the overall platelet range were >0.98. A bias study, however, showed some disagreement. The use of a coincidence correction calculation for the immunoplatelet method did not improve the correlation between the immunoplatelet method and the automated haematology analyser methods. To estimate the possibility of inadequate platelet transfusion, the number of prophylactic platelet transfusion indications determined by the automated haematology analyser platelet counts were compared with the number of transfusion indications according to the platelet counts determined by the immunoplatelet method. An additional 48 blood specimens were included in this analysis. All of the automated haematology analysers showed some disagreement in the transfusion indications when compared with the immunoplatelet method, suggesting the possibility of inadequate platelet transfusion.
机译:将三种自动血液分析仪[Sysmex XE-2100(光学和阻抗模式),Bayer Advia 120和Beckman Coulter LH-750]的准确性与免疫血小板参考方法进行血小板测量进行了比较。从患者中获得了总共165个血液样本,并使用四自动化血液分析仪方法和免疫血小板参考方法确定了血小板计数。全自动血液分析仪方法与免疫血小板参考方法之间的总血小板范围的测定系数(R(2))> 0.98。但是,偏见研究显示出一些分歧。对于免疫血小板方法使用重合校正计算并不能改善免疫血小板方法与自动血液分析仪方法之间的相关性。为了估计血小板输注不充分的可能性,将根据自动血液分析仪的血小板计数确定的预防性血小板输注指征数与根据免疫血小板方法确定的血小板计数的输血指征数进行比较。此分析中还包括48个血液样本。与免疫血小板法相比,所有的自动血液分析仪在输血适应证上均显示出一些分歧,这提示了血小板输注不足的可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号