首页> 外文期刊>International journal of computerized dentistry >Finite element analysis of compressive stress and strain of different implant forms during vertical loading.
【24h】

Finite element analysis of compressive stress and strain of different implant forms during vertical loading.

机译:垂直加载过程中不同植入物形式的压缩应力和应变的有限元分析。

获取原文
获取原文并翻译 | 示例
           

摘要

This study analyzed the influence of platform switching on the stress/strain distribution of vertically loaded implants. Employing finite element analysis, the aim of this study was to determine the distribution of stresses and strains generated by a simulated vertical load on different implant models positioned in the molar region with respect to changes of implant design and biomechanics. Two different implant designs, one with and one without platform switching, with the same length (11.5 mm) and diameter of (3.80 mm) were analyzed. Both virtual implant-bone models were subjected to a load of 500 N. The results showed that the maximum stress/strain in implant-bone samples was located in the area of the implant neck, with stress/strain values decreasing in the apical direction. It was found that the implant design with platform switching developed lower stresses and strains than the implant design without platform switching.
机译:这项研究分析了平台切换对垂直负载植入物的应力/应变分布的影响。利用有限元分析,本研究的目的是确定相对于种植体设计和生物力学的变化,模拟垂直载荷对位于磨牙区的不同种植体模型的模拟垂直载荷产生的应力和应变的分布。分析了两种不同的植入物设计,一种具有和不具有平台切换的设计,它们具有相同的长度(11.5毫米)和直径(3.80毫米)。两种虚拟的植入物骨模型都承受500 N的载荷。结果表明,植入物骨样品中的最大应力/应变位于植入物颈部的区域,应力/应变值在顶端方向减小。发现与不进行平台切换的植入物设计相比,具有平台切换的植入物设计产生的应力和应变更低。