摘要

Periprosthetic fractures can be treated by various methods. The Mennen femoral plate used to be a common implant in our region to stabilise periprosthetic femoral fractures following hip arthroplasty. This device has been used in 16 patients in our region from three different centres. The periprosthetic femoral fractures occurred approximately 7 years after the hip arthroplasty procedure. After stabilisation of the fracture, patients were kept non-weight bearing. Weight bearing was commenced once there was clinical and radiological evidence of fracture union. About 75% of our patients had a complication following fixation of the fracture. In all of these patients the main complication was varus mal-union of the fracture. As per our study the Mennen femoral plate seems to be a weak fixation device. The plate is unable to counter the medial compressive forces on the femur leading to a varus collapse of the fracture.
机译:骨膜周围的骨折可以通过各种方法治疗。 Mennen股骨板曾经是我们地区的常见植入物,用于稳定髋关节置换术后假体周围的股骨骨折。该设备已在我们地区的三个不同中心的16位患者中使用。股骨假体周围骨折发生在髋关节置换术后约7年。骨折稳定后,患者无需承重。一旦有骨折愈合的临床和放射学证据,就开始负重。骨折固定后,约有75%的患者出现并发症。在所有这些患者中,主要并发症是骨折内翻畸形。根据我们的研究,Mennen股骨板似乎是一种弱固定装置。钢板无法抵抗股骨上的内侧压力,从而导致骨折的内翻塌陷。