首页> 外文期刊>Atomic Spectroscopy >An Environmentally Friendly Method for As, Cd, Cr, Cu, Ni, and Pb Determination in Terrestrial Moss Samples Using Ultrasonic Slurry Sampling Combined with Electrothermal Atomic Absorption Spectrometry
【24h】

An Environmentally Friendly Method for As, Cd, Cr, Cu, Ni, and Pb Determination in Terrestrial Moss Samples Using Ultrasonic Slurry Sampling Combined with Electrothermal Atomic Absorption Spectrometry

机译:超声波浆液进样-电热原子吸收光谱法测定陆地藓类样品中的砷,镉,铬,铜,镍和铅

获取原文
获取原文并翻译 | 示例
           

摘要

Novel ultrasonic slurry sampling methods combined with the ETAAS determination of As, Cd, Cr, Cu, Ni, and Pb in terrestrial moss samples were developed without using corrosive acids. A pulverized moss sample between 1 to 50 mg (depending on element concentration) was placed into a 2.5 mL autosampler cup with 2 mL of HNO3 (0.2 M), containing 0.01% of Triton X-100 reagent. The sample was then sonicated directly into the autosampler cup, using an ultrasonic probe at 85% power setting of 7 W for 25 seconds before injection.
机译:开发了一种新型的超声波泥浆采样方法,并结合ETAAS测定陆地苔藓样品中的As,Cd,Cr,Cu,Ni和Pb,而无需使用腐蚀性酸。将1至50 mg(取决于元素浓度)的粉状苔藓样品放入装有2 mL HNO3(0.2 M)的2.5 mL自动进样器杯中,其中含有0.01%的Triton X-100试剂。然后在进样前,使用超声探头以7 W的85%功率设置将样品直接超声处理到自动进样器杯中25秒。