首页> 外文期刊>ASAIO journal >Modular extracorporeal life support: Effects of ultrafiltrate recirculation on the performance of an extracorporeal carbon dioxide removal device
【24h】

Modular extracorporeal life support: Effects of ultrafiltrate recirculation on the performance of an extracorporeal carbon dioxide removal device

机译:模块化体外生命支持:超滤液再循环对体外二氧化碳去除装置性能的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The combination of extracorporeal carbon dioxide removal (ECCO2R) and hemofiltration is a possible therapeutic strategy for patients needing both lung and renal support. We tested the effects of the recirculation of ultrafiltrate on membrane lung (ML) CO2 removal (VCO2ML). Three conscious, spontaneously breathing sheep were connected to a commercially produced ECCO2R device (Hemolung; Alung Technologies, Pittsburgh, PA) with a blood flow of 250 ml/min and a gas flow of 10 L/min. A hemofilter (NxStage, NxStage Medical, Lawrence, MA) was interposed in series after the ML. Ultrafiltrate flow was generated and recirculated before the ML. We tested four ultrafiltrate flows (0, 50, 100, and 150 ml/min) for 25 min each, eight times per animal, resulting in 24 randomized test repetitions. We recorded VCO2ML, hemodynamics and ventilatory variables, and natural lung CO2 transfer (VCO2NL) and collected arterial and circuitry blood samples. VCO2ML was unchanged by application of ultrafiltrate recirculation (40.5 ± 4.0, 39.7 ± 4.2, 39.8 ± 4.2, and 39.2 ± 4.1 ml/min, respectively, at ultrafiltrate flow of 0, 50, 100, and 150 ml/min). Minute ventilation, respiratory rate, VCO2NL, and arterial blood analyses were not affected by ultrafiltrate recirculation. In the tested configuration, ultrafiltrate recirculation did not affect VCO2ML. This modular technology may provide a suitable platform for coupling CO2 removal with various forms of blood purification.
机译:对于需要肺和肾支持的患者,体外二氧化碳去除(ECCO2R)和血液滤过的组合是一种可能的治疗策略。我们测试了超滤液再循环对膜肺(ML)CO2去除(VCO2ML)的影响。将三只自觉呼吸的绵羊连接到商业生产的ECCO2R设备(Hemolung; Alung Technologies,匹兹堡,宾夕法尼亚州),血液流量为250 ml / min,气体流量为10 L / min。 ML后串联插入滤血器(NxStage,NxStage Medical,马萨诸塞州劳伦斯)。在ML之前产生超滤液流并再循环。我们测试了四种超滤液流量(0、50、100和150 ml / min),每次25分钟,每只动物八次,共进行了24次随机测试重复。我们记录了VCO2ML,血液动力学和通气变量以及自然肺CO2转移(VCO2NL),并收集了动脉和循环血样本。通过应用超滤液再循环(分别在0、50、100和150 ml / min的超滤液流速下,分别为40.5±4.0、39.7±4.2、39.8±4.2和39.2±4.1 ml / min),VCO2ML保持不变。分钟通气,呼吸频率,VCO2NL和动脉血分析不受超滤液再循环的影响。在测试的配置中,超滤液再循环不会影响VCO2ML。这种模块化技术可以为将二氧化碳去除与各种形式的血液净化耦合提供合适的平台。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号