首页> 外文期刊>Blood Pressure Monitoring >Extracting oscillometric pulses from the cuff pressure: does it affect the pressures determined by oscillometric blood pressure monitors?
【24h】

Extracting oscillometric pulses from the cuff pressure: does it affect the pressures determined by oscillometric blood pressure monitors?

机译:从袖带压力中提取示波脉冲:它会影响示波血压计确定的压力吗?

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: Oscillometric noninvasive blood pressure measurement devices determine the pressures by analysing the relationship between the pressure in an occluding cuff and low-amplitude pressure pulses (oscillometric pulses) induced in the cuff by the arterial pressure wave. This paper examines the effects on the pulses and oscillometric pulse amplitude envelope of the filters that extract the pulses from the cuff pressure. METHODS: The cuff pressure and oscillometric pulses extracted by the filter were recorded from a Critikon DINAMAP and a Datex Cardiocap, chosen because of accessibility of the filtered and unfiltered signals. The unfiltered oscillometric pulses were determined by subtracting the cuff pressure from a baseline constructed to represent either the Critikon's step deflation cuff pressure or the Datex's gradual pressure decrease. Waveforms were recorded from a noninvasive blood pressure test simulator and three volunteers. CONCLUSIONS: The filter alters the shape of oscillometric pulses causing a shift in the oscillometric pulse amplitude envelope drawn from the filtered pulses compared with that drawn from the unfiltered pulses in human subjects, but not the test simulator. The pulse shape distortion is dependent on the filter characteristics and the oscillometric pulse shape. Further work is required to explore whether this may help explain why simulators with artificial waveforms cannot validate noninvasive blood pressure monitors and why noninvasive blood pressure monitors may not be accurate in all patient groups.
机译:目的:示波法无创血压测量设备通过分析阻塞袖带中的压力与动脉压波在袖带中引起的低振幅压力脉冲(示波脉冲)之间的关系来确定压力。本文研究了对从袖带​​压中提取脉搏的过滤器的脉搏和示波脉冲幅度包络的影响。方法:从Critikon DINAMAP和Datex Cardiocap记录由过滤器提取的袖带压力和示波脉冲脉冲,这是由于可过滤和未过滤信号的可及性而选择的。未过滤的示波脉冲是通过从构造为代表Critikon阶状放气袖带压或Datex逐渐压降的基线中减去袖带压来确定的。记录了无创血压测试模拟器和三名志愿者的波形。结论:滤波器改变了示波脉冲的形状,从而使从被滤波的脉冲中抽出的示波脉冲幅度包络与从人类受试者中未滤波的脉冲中抽取出的示波脉冲幅度包络发生了变化,但测试模拟器却没有。脉冲形状失真取决于滤波器特性和示波脉冲形状。需要做进一步的工作来探索这是否可以帮助解释为什么带有人工波形的模拟器无法验证无创血压监测器,以及为什么无创血压监测器可能在所有患者组中都不准确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号