A novel quartz device has been designed to trap arsine and selenium hydride and subsequently to volatilize the collected analyte and atomize it for atomic-absorption spectrometric detection. The device is actually the multiple microflame quartz-tube atomizer (multiatomizer) with inlet arm modified to serve as the trap and to accommodate the oxygen-delivery capillary used to combust hydrogen during the trapping step. The effect of relevant experimental conditions (trap temperature during trapping and hydrogen flow rate and trap temperature during volatilization) on collection and volatilization efficiency was investigated. Under the optimum conditions collection and volatilization efficiency for arsenic and selenium were 50 and 70%, respectively.
机译:设计了一种新颖的石英装置,以捕集a和氢化硒,随后挥发收集的分析物并将其雾化以用于原子吸收光谱检测。该设备实际上是多微火焰石英管雾化器(多雾化器),其进口臂经过了改进,可以用作捕集器并容纳用于在捕集步骤中燃烧氢气的氧气输送毛细管。研究了相关实验条件(捕集期间的捕集温度和挥发期间的氢气流速以及挥发期间的捕集温度)对收集和挥发效率的影响。在最佳条件下,砷和硒的收集和挥发效率分别为50%和70%。