An improved apparatus for on-line chemiluminescence detection with capillary electrophoresis and detection of metal-porphyrins

机译:一种改进的毛细管电泳在线化学发光检测和金属卟啉检测装置

获取原文
获取原文并翻译 | 示例
           

摘要

An improved apparatus for on-line chemiluminescence detection with capillary electrophoresis is proposed. In the present method the grounding electrode of the power was put in the four-way joint, not in the eluate reservoir, and the reaction capillary with greater inner diameter was set at a slope down. The bubbles produced by the breakdown of hydrogen peroxide could therefore be drained out much more easily and did not perturb the separation and detection. Satisfactory reproducibility was still achieved even at a higher concentration of hydrogen peroxide in strongly basic medium. Metal-porphyrins, the sensitive catalysts for luminol chemiluminescence reactions, were measured for the first time by on-line chemiluminescence detection with capillary electrophoresis. The detection limit for 5,10,15,20-terakis(4-sulfonatophenyl)-21H,23H-porphine manganese(III) was 361 amol and 8.1 * 10~(-9) mol L~(-1). Under the optimized chemiluminescence conditions. Co-meso-tetrakis(4-sulfonatophenyl)porphyrin with limited catalytic efficiency could also be detected with a satisfactory detection limit of 964 fmol and 2.7 * 10~(-5) mol L~(-1).
机译:提出了一种改进的毛细管电泳在线化学发光检测装置。在本方法中,将电源的接地电极放在四通接头中,而不是在洗脱液容器中,并将内径较大的反应毛细管向下倾斜。因此,由过氧化氢分解产生的气泡可以更容易地排出,并且不会干扰分离和检测。即使在强碱性介质中以较高的过氧化氢浓度仍可实现令人满意的重现性。金属卟啉是用于鲁米诺化学发光反应的敏感催化剂,首次通过毛细管电泳在线化学发光检测进行了首次测量。 5,10,15,20-叔(4-磺酰基苯基)-21H,23H-吗啡锰(III)的检出限为361 amol和8.1 * 10〜(-9)mol L〜(-1)。在优化的化学发光条件下。还可以检测到催化效率有限的间-间四(4-磺酰基苯基)卟啉,其满意的检出限为964 fmol和2.7 * 10〜(-5)mol L〜(-1)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号