首页> 外文期刊>Biomedical Chromatography: An International Journal Devoted to Research in Chromatographic Methodologies and Their Applications in the Biosciences >Simultaneous detection of fluquinconazole and flusilazole in lettuce using gas chromatography with a nitrogen phosphorus detector: decline patterns at two different locations
【24h】

Simultaneous detection of fluquinconazole and flusilazole in lettuce using gas chromatography with a nitrogen phosphorus detector: decline patterns at two different locations

机译:使用气相色谱仪和氮磷检测器同时检测生菜中的氟喹康唑和氟硅唑:在两个不同位置的下降模式

获取原文
获取原文并翻译 | 示例
           

摘要

Method validations in addition to decline patterns of fluquinconazole and flusilazole in lettuce grown under greenhouse conditions at two different locations were investigated. Following the application of fluquinconazole and flusilazole at a dose rate of 20mL/20L water, lettuce samples were collected randomly for up to 7days post-application, and simultaneously extracted with acetone, purified through solid-phase extraction, analyzed via gas chromatography with a nitrogen phosphorus detector, and confirmed through gas chromatography-mass spectrometry. The linearity was excellent, with determination coefficients (R-2) between 0.9999 and 1.0. The method was validated in triplicate at two different spiking levels (0.2 and 1.0mg/kg) with satisfactory recoveries between 75.7 and 97.9% and relative standard deviations of <9. The limit of quantification was 0.01mg/kg. Both analytes declined very quickly, as can be seen from the short half-life time of <4days. Statistical analysis revealed significant differences between residues at different days of sampling, except at 7days post-application (triple application). At that point, the decline patterns of fluquinconazole and flusilazole were independent of application rate, location, temperature and humidity. Copyright (c) 2015 John Wiley & Sons, Ltd.
机译:除了在温室条件下在两个不同地点种植的生菜中氟喹康唑和氟硅唑的下降模式外,还对方法的有效性进行了研究。在以20mL / 20L水的剂量比例施用氟喹康唑和氟硅唑之后,在施用后长达7天的时间内随机收集生菜样品,同时用丙酮萃取,通过固相萃取纯化,通过氮气色谱法进行分析磷检测器,并通过气相色谱-质谱法确认。线性非常好,测定系数(R-2)在0.9999和1.0之间。该方法在两个不同的加标水平(0.2和1.0mg / kg)下重复三次进行验证,回收率在75.7%至97.9%之间,相对标准偏差<9。定量限为0.01mg / kg。从小于4天的短半衰期可以看出,两种分析物的下降速度都非常快。统计分析表明,除了在施用后7天(三重施用)外,在不同采样天的残留物之间存在显着差异。那时,氟喹康唑和氟硅唑的下降模式与施用量,位置,温度和湿度无关。版权所有(c)2015 John Wiley&Sons,Ltd.

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号