? PURPOSE: To assess through-focus polychromatic image sharpness of 5 FDA-approved presbyopia-correcting intraocular lenses (IOLs) through a range of object vergences and pupil diameters using an image sharpness algorithm. ? DESIGN: Laboratory investigation. ? METHODS: A 1951 USAF resolution target was imaged through Crystalens AO (AO), Crystalens HD (HD), aspheric ReSTOR +4 (R4), aspheric ReSTOR +3 (R3), and Tecnis Multifocal Acrylic (TMF) IOL in a model eye and captured digitally for each combination of pupil diameter and object vergence. The sharpness of each digital image was objectively scored using a 2-dimensional gradient function. ? RESULTS: AO had the best distance image sharpness for all pupil diameters and was superior to the HD. With a 5-mm pupil, the R4 distance image sharpness was similar to the HD and at 6 mm the TMF was superior to the HD, R3, and R4. The R3 moved the near focal point farther from the patient compared to the R4, but did not improve image sharpness at intermediate distances and showed worse distance and near image sharpness. Consistent with apodization, the ReSTOR IOLs displayed better distance and poorer near image sharpness as pupil diameter increased. TMF showed consistent distance and near image sharpness across pupil diameters and the best near image sharpness for all pupil diameters. ? CONCLUSIONS: Differing IOL design strategies to increase depth of field are associated with quantifiable differences in image sharpness at varying vergences and pupil sizes. Objective comparison of the imaging properties of specific presbyopia-correcting IOLs in relation to patient's pupil sizes can be useful in selecting the most appropriate IOL for each patient.
机译:?目的:使用图像清晰度算法,通过一系列的物体散度和瞳孔直径,评估5种FDA批准的老花眼矫正人工晶状体(IOL)的全焦点多色图像清晰度。 ?设计:实验室调查。 ?方法:在模型眼中,通过Crystalens AO(AO),Crystalens HD(HD),非球面ReSTOR +4(R4),非球面ReSTOR +3(R3)和Tecnis多焦点丙烯酸(TMF)IOL对1951 USAF分辨率目标成像并针对瞳孔直径和物体发散度的每种组合进行数字捕获。使用二维梯度函数客观地评分每个数字图像的清晰度。 ?结果:对于所有瞳孔直径,AO都具有最佳的距离图像清晰度,并且优于HD。对于5mm的瞳孔,R4距离图像的清晰度与HD相似,而在6mm处,TMF优于HD,R3和R4。与R4相比,R3将近焦点移离患者更远,但在中间距离处图像清晰度没有改善,并且距离和近图像清晰度更差。与变迹相一致,随着瞳孔直径的增加,ReSTOR IOL显示出更好的距离和更差的近图像清晰度。 TMF在整个瞳孔直径上显示出一致的距离和近图像锐度,并且在所有瞳孔直径下都具有最佳的近图像锐度。 ?结论:增加景深的不同IOL设计策略与在不同散度和瞳孔大小下图像清晰度的可量化差异相关。将特定的老花眼矫正IOL的成像特性与患者瞳孔大小相关的客观比较有助于为每个患者选择最合适的IOL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号